Biomechanical performance of subpectoral biceps tenodesis: a comparison of interference screw fixation, cortical button fixation, and interference screw diameter.

机构信息

ONS Foundation for Clinical Research and Education, ONS, Greenwich, CT, USA.

出版信息

J Shoulder Elbow Surg. 2013 Apr;22(4):451-7. doi: 10.1016/j.jse.2012.03.016. Epub 2012 Jun 26.

Abstract

BACKGROUND

Subpectoral biceps tenodesis with interference screw fixation allows reproducible positioning of the tendon to help maintain the length-tension relationship. The aim of our study was to evaluate the role of cortical button fixation in isolation or as an augment to interference screw fixation and to determine if the diameter of the interference screw affected fixation strength.

MATERIALS AND METHODS

Thirty-two cadaveric shoulders were dissected and randomized to 1 of 4 groups: (1) 7-mm interference screw and cortical button, (2) cortical button alone, (3) 7-mm interference screw, or (4) 8-mm interference screw. Testing was performed on a materials testing system with a 100-N load cycled at 1 Hz for 5000 cycles, followed by an axial load to failure test. Cyclic displacement, ultimate load to failure, and site of failure were recorded for each specimen.

RESULTS

The mean ultimate failure loads were 7-mm interference screw with cortical button augmentation, 237.8 ± 120.4 N; cortical button alone, 99.4 ± 16.9 N; 7-mm interference screw, 275.5 ± 56 N; 8-mm interference screw, 277.1 ± 42.1 N. All specimens failed through tendon failure at the screw-tendon-bone interface.

CONCLUSIONS

The biomechanical performance of subpectoral biceps tenodesis with interference screw fixation was not improved with cortical button augmentation. In addition, cortical button fixation alone yielded a significantly lower ultimate load to failure compared with interference screws. Finally, the biomechanical performance of smaller-diameter interference screws with matching bone tunnels was not affected by interference screw diameter.

摘要

背景

带皮质骨纽扣的下肌止点肱二头肌固定术可使肌腱的定位具有可重复性,有助于维持长度-张力关系。我们的研究目的是评估皮质骨纽扣固定术在单纯使用或作为干扰螺钉固定术的辅助手段时的作用,并确定干扰螺钉的直径是否会影响固定强度。

材料与方法

对 32 具尸体肩关节进行解剖并随机分为 4 组:(1)7mm 干扰螺钉和皮质骨纽扣,(2)皮质骨纽扣单独使用,(3)7mm 干扰螺钉,或(4)8mm 干扰螺钉。测试在材料测试系统上进行,以 1Hz 的频率循环 100N 载荷 5000 次,随后进行轴向失效试验。记录每个标本的循环位移、失效极限载荷和失效部位。

结果

7mm 干扰螺钉加皮质骨纽扣增强组的平均失效极限载荷为 237.8±120.4N;皮质骨纽扣单独使用组为 99.4±16.9N;7mm 干扰螺钉组为 275.5±56N;8mm 干扰螺钉组为 277.1±42.1N。所有标本均在螺钉-肌腱-骨界面处发生肌腱失效。

结论

带皮质骨纽扣的下肌止点肱二头肌固定术的生物力学性能并未因皮质骨纽扣增强而提高。此外,皮质骨纽扣单独固定的失效极限载荷明显低于干扰螺钉。最后,与匹配的骨隧道相比,较小直径的干扰螺钉的生物力学性能不受干扰螺钉直径的影响。