Ordered subset expectation maximisation with depth-dependent resolution recovery (OSEM-RR) is a processing algorithm reported to improve images with halved tracer activity in myocardial perfusion scintigraphy (MPS) compared to filtered backprojection (FBP) using conventional activities. OSEM-RR has not yet been compared with maximal likelihood expectation maximisation (MLEM).
39 patients undergoing MPS and two anthropomorphic phantoms (one with, one without an inferior wall insert) had full-time (FT) and half-time (HT) SPECT datasets acquired simultaneously and processed by FBP, MLEM and OSEM-RR. Two experienced reporters scored images of all clinical studies (n=234) for conspicuity of a perfusion defect, with results being compared using Wilcoxon paired and Kappa tests. A quantitative assessment based on mean segmental pixel counts taken from numbers automatically displayed over the 20 segments of Cedars Sinai Autoquant QPS image were compared using Pearson's correlation and Bland Altman analysis.
A small but consistent superior concurrence between FT and HT datasets for OSEM-RR compared to FBP and MLEM was observed for both qualitative and quantitative analyses. OSEM-RR resulted in better definition of the inferior wall defect on the phantom study.
OSEM-RR appears superior to both FBP and MLEM in terms of handling reduced count statistics.
与使用传统放射性活度的滤波反投影(filtered backprojection,FBP)相比,基于深度依赖分辨率恢复(depth-dependent resolution recovery,RR)的有序子集期望最大化(ordered subset expectation maximisation,OSEM)处理算法可改善心肌灌注闪烁照相术(myocardial perfusion scintigraphy,MPS)中放射性示踪剂活度减半的图像。目前尚未将 OSEM-RR 与最大似然期望最大化(maximal likelihood expectation maximisation,MLEM)进行比较。
39 名接受 MPS 检查的患者和 2 个人体模型(一个模型带有,一个不带下壁插入物)同时采集全时(full-time,FT)和半时(half-time,HT)SPECT 数据集,并分别使用 FBP、MLEM 和 OSEM-RR 进行处理。2 位经验丰富的报告员对所有临床研究(n=234)的图像进行了显影缺陷的可视性评分,并使用 Wilcoxon 配对和 Kappa 检验进行比较。基于 Cedars Sinai Autoquant QPS 图像上自动显示的 20 个节段的像素计数平均值,使用 Pearson 相关和 Bland Altman 分析进行了定量评估。
在定性和定量分析中,与 FBP 和 MLEM 相比,OSEM-RR 观察到 FT 和 HT 数据集之间的一致性略有提高。OSEM-RR 可改善人体模型研究中下壁缺陷的定义。
OSEM-RR 在处理计数统计数据减少方面似乎优于 FBP 和 MLEM。