Anal Biochem. 2012 Oct 1;429(1):58-69. doi: 10.1016/j.ab.2012.06.024. Epub 2012 Jul 2.
To characterize a proprietary therapeutic monoclonal antibody (mAb) candidate, a rigorous biophysical study consisting of 53 Biacore and kinetic exclusion assay (KinExA) experiments was undertaken on the therapeutic mAb complexing with its target antigen. Unexpectedly, the observed binding kinetics depended on the chip used, suggesting that the negatively charged carboxyl groups on CM5, CM4, and C1 chips were adversely affecting the Biacore kinetic results. To study this hypothesis, Biacore solution-phase and KinExA equilibrium titrations, as well as KinExA kinetic measurements, were performed to establish accurate values for the affinity and kinetic rate constants of the binding reaction between antigen and mAb. The results revealed that as the negative charge on the biosensor surface decreased, the binding kinetics and K(D) approached the accurate binding parameters more closely when measured in solution. Two potential causes for the artifactual Biacore surface-based measurements are (i) steric hindrance of antigen binding arising from an interaction of the negatively charged carboxymethyldextran matrix with the mAb, which is a highly basic protein with a pI of 9.4, and (ii) an electrostatic repulsion between the negatively charged antigen and the carboxymethyldextran matrix. Importantly, simple diagnostic tests can be performed early in the measurement process to identify these types of matrix-mediated artifacts.
为了对一种专有的治疗性单克隆抗体(mAb)候选物进行表征,对该治疗性 mAb 与靶抗原的复合物进行了一项严格的生物物理研究,其中包括 53 个 Biacore 和动力学排除测定(KinExA)实验。出乎意料的是,观察到的结合动力学取决于所用的芯片,这表明 CM5、CM4 和 C1 芯片上带负电荷的羧基基团对 Biacore 动力学结果产生了不利影响。为了研究这一假设,进行了 Biacore 溶液相和 KinExA 平衡滴定以及 KinExA 动力学测量,以确定抗原与 mAb 之间结合反应的亲和力和动力学速率常数的准确值。结果表明,随着生物传感器表面上负电荷的减少,当在溶液中测量时,结合动力学和 K(D)更接近准确的结合参数。Biacore 表面基测量出现人为假象的两个潜在原因是:(i) 带负电荷的羧甲基葡聚糖基质与 mAb 相互作用,导致抗原结合的空间位阻,而 mAb 是一种具有等电点(pI)为 9.4 的高度碱性蛋白;(ii) 带负电荷的抗原与羧甲基葡聚糖基质之间的静电排斥。重要的是,可以在测量过程的早期进行简单的诊断测试,以识别这些类型的基质介导的假象。