Estimation of absorbed organ doses and effective dose based on body mass index in digital radiography.

作者信息

Kim Hyunji, Park MinSeok, Park Songyi, Jeong Hoiwoun, Kim Jungmin, Kim Youhyun

机构信息

Department of Radiologic Science, College of Health Science, Korea University, Seoul, Republic of Korea.

出版信息

Radiat Prot Dosimetry. 2013 Jan;153(1):92-9. doi: 10.1093/rpd/ncs089. Epub 2012 Jul 6.

Abstract

With the introduction of digital radiography, patients undergoing radiographic procedures are subject to being overexposed to radiation. Therefore, it is necessary to estimate the absorbed organ dose and the effective dose, which are significant for patient health, along with body type. During chest radiographic examinations conducted in 899 patients for screening, the absorbed dose of the 13 major organs, the average whole-body dose, and two effective doses weighted by factors published in ICRP 60 and ICRP 103 were calculated on the basis of patient information such as height, weight and examination condition, including kilovolt potential, focus-skin distance and entrance surface dose (ESD), using a PC-based Monte Carlo program simulation. It was found that dose per unit ESD had a tendency to decrease with body mass index (BMI). In particular, the absorbed dose for most organs was larger at high voltages (140 kVp) than at low voltages (120 kVp, 100 kVp). In addition, the effective dose which was based on ICRP 60 and ICRP 103 also represented the same tendency in respect of BMI and tube voltage.