Miyamoto Y, Kormos R L, Borovetz H S, Gaisor T, Pristas J M, Armitage J M, Hardesty R L, Griffith B P
Department of Surgery, School of Medicine, University of Pittsburgh, Pennsylvania 15261.
Artif Organs. 1990 Dec;14(6):454-7. doi: 10.1111/j.1525-1594.1990.tb03003.x.
The interrelationships between hemodynamic variables including right ventricular (RV) performance with filling/ejection dynamics of the Novacor left ventricular assist system (LVAS) were determined in 10 of 11 patients who received LVAS as a bridge to heart transplant. Nine were successfully transplanted. Data were obtained intraoperatively, at periodic intervals up to 48 h postimplant and at explant. The hypotheses investigated included (a) RV performance influences LVAS filling characteristics and (b) LVAS pump output is influenced by systemic vascular resistance (SVR). During the period of LVAS support (2-126 days), pumping characteristics included a mean filling volume of 51 ml (range, 24-70), residual volume of 4.9 ml (range, 1-18), pump rate of 113/min (range, 63-175), and pump output of 5.81/min (range, 2.8-8.2). Multiple regression analysis identified pulmonary vascular resistance (PVR), RV stroke work index (RVSWI), and pulmonary capillary wedge pressure, but not RV ejection fraction, pulmonary artery pressure, or central venous pressure (CVP) as the most important correlates with LVAS filling volume (p less than 0.001, R2 = 0.6). In addition, LVAS pump output was influenced mainly by RVSWI, PVR, and SVR (p less than 0.001, R2 = 0.7). It was concluded that LVAS performance is highly dependent on RV function and systemic/pulmonary vascular resistances.
在11例接受诺瓦科尔左心室辅助系统(LVAS)作为心脏移植过渡治疗的患者中,对其中10例患者的血流动力学变量之间的相互关系进行了研究,这些变量包括右心室(RV)功能以及LVAS的充盈/射血动力学。9例患者成功接受了心脏移植。在术中、植入后长达48小时的定期时间点以及植入物取出时获取数据。所研究的假设包括:(a)右心室功能影响LVAS的充盈特征;(b)LVAS的泵输出受全身血管阻力(SVR)影响。在LVAS支持期间(2 - 126天),泵的特征包括平均充盈量为51毫升(范围为24 - 70毫升)、残余量为4.9毫升(范围为1 - 18毫升)、泵速为113次/分钟(范围为63 - 175次/分钟)以及泵输出为5.8升/分钟(范围为2.8 - 8.2升/分钟)。多元回归分析确定肺血管阻力(PVR)、右心室每搏功指数(RVSWI)和肺毛细血管楔压,但不包括右心室射血分数、肺动脉压或中心静脉压(CVP)是与LVAS充盈量最相关的因素(p < 0.001,R2 = 0.6)。此外,LVAS的泵输出主要受RVSWI、PVR和SVR影响(p < 0.001,R2 = 0.7)。研究得出结论,LVAS的性能高度依赖于右心室功能以及全身/肺血管阻力。