An easily performed reference haematocrit would be useful in quality control, in the preparation of whole-blood calibrators and in the evaluation of automated, whole-blood analysers. Available techniques are complex, time-consuming and require radiolabelled reagents that are difficult to obtain. They have largely been directed towards correction of 'trapped plasma' errors. Other errors which affect haematocrit measurement but which have not yet been addressed include those due to white-cell contamination, those due to defects in tube plugging and those resulting from red-cell dehydration. We propose a simple, rapid and reliable reference haematocrit procedure which relies on the fact that haematocrit equals haemoglobin/MCHC and that MCHC determinations via haemoglobin analyses can be performed accurately on as little as 6 microliters of packed red cells (CV less than 0.5%). Selection of the cells to be analysed from the mid portion of a packed red-cell column eliminates interference from white-cell contamination/red-cell dehydration. The absence of trapped plasma in this portion of the red cell column after centrifugation for 4 min at 10,000 r.c.f. can be documented by a second cycle of centrifugation.
一种易于操作的参考血细胞比容在质量控制、全血校准品的制备以及自动全血分析仪的评估中会很有用。现有的技术复杂、耗时,并且需要难以获得的放射性标记试剂。它们主要致力于校正“滞留血浆”误差。其他影响血细胞比容测量但尚未得到解决的误差包括由于白细胞污染、试管堵塞缺陷以及红细胞脱水导致的误差。我们提出了一种简单、快速且可靠的参考血细胞比容程序,该程序基于血细胞比容等于血红蛋白/MCHC这一事实,并且通过血红蛋白分析对MCHC的测定可以在低至6微升的压积红细胞上准确进行(变异系数小于0.5%)。从压积红细胞柱中部选择待分析的细胞可消除白细胞污染/红细胞脱水的干扰。在10,000相对离心力下离心4分钟后,红细胞柱这一部分不存在滞留血浆可通过第二次离心循环来证明。