We describe a previously unreported phenomenon of intermittent outflow right ventricular tract capture from the atrial lead of a dual-chamber pacemaker. This was more obvious at slower paced atrial rates and disappeared by decreasing the atrial pulses voltage. Electroanatomical mapping showed that the onset of activation was nearly simultaneous at the insertion site of the atrial lead and at an intermediate level of the right ventricular outflow tract. This exceptional finding might be erroneously diagnosed as due to pseudo-pseudo fusion beats.
我们描述了一种先前未报告的现象,即双腔起搏器心房电极间歇性夺获右心室流出道。在较慢的心房起搏频率下这种现象更明显,通过降低心房脉冲电压可使其消失。电解剖标测显示,激动起始在心房电极插入部位和右心室流出道中间水平几乎同时发生。这一特殊发现可能会被错误地诊断为伪伪融合波所致。