Huang Yen-Cho, Yeh Chien-Yi, Yeh Jih-Hsiang, Lo Ching-Jung, Tsai Ping-Fang, Hung Chih-Hung, Tsai Chieh-Sheng, Chen Chen-Yuan
Department of Radiation Oncology, Chang Gung Memorial Hospital, Keelung Branch, Keelung City, Taiwan.
Med Dosim. 2013 Spring;38(1):35-41. doi: 10.1016/j.meddos.2012.05.004. Epub 2012 Jul 31.
Volumetric-modulated arc therapy (VMAT) is a novel extension of the intensity-modulated radiation therapy (IMRT) technique, which has brought challenges to dose verification. To perform VMAT pretreatment quality assurance, an electronic portal imaging device (EPID) can be applied. This study's aim was to evaluate EPID performance for VMAT dose verification. First, dosimetric characteristics of EPID were investigated. Then 10 selected VMAT dose plans were measured by EPID with the rotational method. The overall variation of EPID dosimetric characteristics was within 1.4% for VMAT. The film system serving as a conventional tool for verification showed good agreement both with EPID measurements ([94.1 ± 1.5]% with 3 mm/3% criteria) and treatment planning system (TPS) calculations ([97.4 ± 2.8]% with 3 mm/3% criteria). In addition, EPID measurements for VMAT presented good agreement with TPS calculations ([99.1 ± 0.6]% with 3 mm/3% criteria). The EPID system performed the robustness of potential error findings in TPS calculations and the delivery system. This study demonstrated that an EPID system can be used as a reliable and efficient quality assurance tool for VMAT dose verification.
容积调强弧形放疗(VMAT)是调强放射治疗(IMRT)技术的一种新扩展,给剂量验证带来了挑战。为进行VMAT治疗前质量保证,可应用电子射野影像装置(EPID)。本研究的目的是评估EPID在VMAT剂量验证方面的性能。首先,研究了EPID的剂量学特性。然后采用旋转法通过EPID测量了10个选定的VMAT剂量计划。对于VMAT,EPID剂量学特性的总体变化在1.4%以内。作为传统验证工具的胶片系统与EPID测量结果(3 mm/3%标准下为[94.1±1.5]%)和治疗计划系统(TPS)计算结果(3 mm/3%标准下为[97.4±2.8]%)均显示出良好的一致性。此外,VMAT的EPID测量结果与TPS计算结果(3 mm/3%标准下为[99.1±0.6]%)也显示出良好的一致性。EPID系统在TPS计算和输送系统中表现出发现潜在误差的稳健性。本研究表明,EPID系统可作为VMAT剂量验证的可靠且高效的质量保证工具。