Hata Junichi, Yagi Kazuo, Hikishima Keigo, Numano Tomokazu, Goto Masami, Yano Keichi
Department of Radiology, University of Tokyo Hospital, 7-3-1 Hongoh, Bunkyo ward, Tokyo, 113-8655 Japan.
Radiol Phys Technol. 2013 Jan;6(1):92-7. doi: 10.1007/s12194-012-0174-1. Epub 2012 Aug 15.
The aim of our study was to simulate an effective diffusion-weighted imaging (DWI) pulse sequence and to evaluate the best b value in skeletal muscle. The evaluated pulse sequences were spin echo (SE), stimulated echo (STE), and gradient-recalled echo (GRE). The signal intensity changed in some DWI pulse sequences when the b value was changed. Moreover, the static magnetic field of 1.5 T was compared with that of 3.0 T. Next, diffusion tensor imaging of the human skeletal muscle was measured. The experimental subjects were healthy male volunteers (n = 7; age 21.8 ± 1.1 years). We changed the b value in steps of 100 s/mm(2) from 0 to 2000 s/mm(2), and the diffusion values [fractional anisotropy (FA), λ(1), λ(2), and λ(3)] were calculated based on the data. The STE method could maintain signals up to b values of 1300 s/mm(2), but the SE and GRE methods suffered from high noise. In the human study, the mean FA (±SD) was 0.41 ± 0.02 in the tibialis anterior muscle (TA) and 0.31 ± 0.02 in the soleus muscle (SOL) at a b value of 1000 s/mm(2). The correlation of the FA with the b value coefficient and the P value by Student's t test were r = 0.981 and P < 0.001 for the TA and r = 0.982, P < 0.001 for the SOL. However, the FA became high with a low b value. In conclusion, STE was very useful for DWI in a short T (2)-value tissue. Moreover, a minimum b value of 800 s/mm(2) was necessary for evaluation of human skeletal muscle.
我们研究的目的是模拟一种有效的扩散加权成像(DWI)脉冲序列,并评估骨骼肌中的最佳b值。评估的脉冲序列有自旋回波(SE)、受激回波(STE)和梯度回波(GRE)。当b值改变时,一些DWI脉冲序列中的信号强度会发生变化。此外,还比较了1.5 T和3.0 T的静磁场。接下来,对人体骨骼肌进行了扩散张量成像测量。实验对象为健康男性志愿者(n = 7;年龄21.8±1.1岁)。我们将b值以100 s/mm²为步长从0改变到2000 s/mm²,并根据数据计算扩散值[分数各向异性(FA)、λ(1)、λ(2)和λ(3)]。STE方法在b值达到1300 s/mm²时仍能保持信号,但SE和GRE方法存在高噪声。在人体研究中,当b值为1000 s/mm²时,胫骨前肌(TA)的平均FA(±标准差)为0.41±0.02,比目鱼肌(SOL)为0.31±0.02。通过学生t检验,TA的FA与b值系数的相关性及P值分别为r = 0.981和P < 0.001,SOL的r = 0.982,P < 0.001。然而,低b值时FA会变高。总之,STE在短T(2)值组织的DWI中非常有用。此外,评估人体骨骼肌需要的最小b值为800 s/mm²。