Brockett Claire L, Jennings Louise M, Hardaker Catherine, Fisher John
Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds, UK.
Proc Inst Mech Eng H. 2012 Jul;226(7):529-35. doi: 10.1177/0954411912445265.
Cross-linked polyethylene has been introduced into total joint replacement to improve wear resistance. Although the performance of highly cross-linked polyethylene is well documented clinically and experimentally for total hip replacements, the reduction in mechanical properties with increasing irradiation is of concern for application to total knee replacement. The aim of this study was to investigate the wear performance of a moderately cross-linked polyethylene material in a fixed-bearing total knee replacement. The study was conducted using two femoral geometries, a conventional cruciate-retaining femoral and a high-flexion femoral geometry. The femoral geometry appeared to have no effect on the wear of the knee replacement under standard gait conditions. A significant reduction in wear volume was measured with the moderately cross-linked polyethylene compared with the conventional polyethylene over a six-million-cycle wear study. This study indicates the use of a moderately cross-linked polyethylene in a fixed-bearing total knee replacement may provide a low wearing option for total knee replacement.
交联聚乙烯已被引入全关节置换术以提高耐磨性。尽管高度交联聚乙烯在全髋关节置换术的临床和实验性能已有充分记录,但随着辐照增加机械性能的降低对于应用于全膝关节置换术而言令人担忧。本研究的目的是研究一种适度交联聚乙烯材料在固定平台全膝关节置换术中的磨损性能。该研究使用了两种股骨几何形状,一种是传统的保留交叉韧带的股骨几何形状和一种高屈曲股骨几何形状。在标准步态条件下,股骨几何形状似乎对膝关节置换术的磨损没有影响。在一项六百万次循环的磨损研究中,与传统聚乙烯相比,适度交联聚乙烯的磨损体积显著减少。这项研究表明,在固定平台全膝关节置换术中使用适度交联聚乙烯可能为全膝关节置换术提供一种低磨损的选择。