Department of Radiology, Northwestern University, Chicago, IL 60611, USA.
J Vasc Interv Radiol. 2012 Sep;23(9):1233-5. doi: 10.1016/j.jvir.2012.06.001.
Various adjunctive techniques have been reported for challenging inferior vena cava (IVC) filter retrievals (1-4). One particularly challenging obstacle to successful IVC filter retrieval is the formation of a radiolucent fibrin cap over the embedded apex of the IVC filter. This obstacle is a result of filter tilting that creates turbulent blood flow, which promotes fibrin cap formation. The aim of this brief report is to depict a novel technique for IVC filter retrieval: guide wire-manipulated disruption of the fibrin cap.
各种辅助技术已被报道用于解决下腔静脉(IVC)滤器取出的难题(1-4)。在 IVC 滤器取出成功的过程中,一个特别具有挑战性的障碍是嵌入 IVC 滤器尖端的不透明纤维蛋白帽的形成。这个障碍是由于滤器倾斜造成的湍流血流,促进了纤维蛋白帽的形成。本简要报告的目的是描述一种新的 IVC 滤器取出技术:导丝操作破坏纤维蛋白帽。