Suppr超能文献

清醒兔心输出量测量的多普勒方法:验证研究、用途及局限性

The Doppler method for measuring cardiac output in conscious rabbits: validation studies, uses, and limitations.

作者信息

Hof R P, Hof A, Stürm R P

机构信息

Cardiovascular Unit of Preclinical Research, Sandoz Ltd., Basel, Switzerland.

出版信息

J Pharmacol Methods. 1990 Dec;24(4):263-76. doi: 10.1016/0160-5402(90)90011-9.

Abstract

The velocity of blood in the rabbit aorta is very fast, approaching the limits of some pulsed directional Doppler flow meters. Therefore, we thoroughly evaluated a 20- and a 10-MHz device for measuring cardiac output in rabbits. Flow probes were implanted around the ascending aorta and catheters were implanted into the left atrium (for microsphere injection), femoral artery, and vein. About 2 weeks later, cardiac output was determined with the Doppler method and simultaneously with tracer microspheres. Cardiac output was manipulated with isoproterenol, dihydralazine, guanfacine, or alinidine, intravenously. With the 20-MHz device, only normal and decreased cardiac output could be measured accurately, even with a 60 degrees implantation angle of the crystals. The 10-MHz device yielded accurate measurements also at very high flow. Surprisingly, at high aortic flow rates, both the 10- and the 20-MHz devices were unable to measure correctly diastolic flow, which is close to zero. It was necessary to adjust the position of the late diastolic Doppler signal manually to the electrical zero line. With this precaution, the 10-MHz device yielded an excellent correlation between mean Doppler signal and cardiac output. Cardiac output can be measured in absolute flow units if the flow-probe can be calibrated in vivo with an independent, accurate method about 2 wk after implantation. The baroreflex was not affected by the implanted flow probes. Within these limits, Doppler flow meters are good tools to assess drug effects on cardiac output in conscious rabbits.

摘要

兔子主动脉中的血流速度非常快,接近一些脉冲式定向多普勒流量计的测量极限。因此,我们全面评估了一款20兆赫兹和一款10兆赫兹的设备用于测量兔子的心输出量。在升主动脉周围植入流量探头,并将导管植入左心房(用于微球注射)、股动脉和静脉。大约2周后,用多普勒方法并同时用示踪微球测定心输出量。通过静脉注射异丙肾上腺素、双肼屈嗪、胍法辛或阿利尼定来控制心输出量。使用20兆赫兹的设备,即使晶体植入角度为60度,也只能准确测量正常和降低的心输出量。10兆赫兹的设备在非常高的流量下也能进行准确测量。令人惊讶的是,在高主动脉流速下,10兆赫兹和20兆赫兹的设备都无法正确测量接近零的舒张期血流。有必要手动将舒张期末期多普勒信号的位置调整到电零线。采取这一预防措施后,10兆赫兹的设备在平均多普勒信号和心输出量之间产生了极好的相关性。如果流量探头在植入后约2周能够用一种独立、准确的方法在体内进行校准,心输出量可以用绝对流量单位来测量。压力反射不受植入的流量探头影响。在这些限制范围内,多普勒流量计是评估药物对清醒兔子心输出量影响的良好工具。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验