一种用于连续监测心输出量变化的可移动腔外多普勒探头。

A removable extraluminal Doppler probe for continuous monitoring of changes in cardiac output.

作者信息

Keagy B A, Lucas C L, Hsiao H S, Wilcox B R

出版信息

J Ultrasound Med. 1983 Aug;2(8):357-62. doi: 10.7863/jum.1983.2.8.357.

Abstract

To address the problem of monitoring cardiac output in postoperative cardiac patients, a removable, extraluminal pulsed Doppler probe has been constructed which continuously reflects changes in cardiac output by monitoring blood velocity in the ascending aorta. Velocity is related to the frequency of the Doppler shift and the angle at which the ultrasound beam intersects the vessel. A 1 mm2 piezoelectric crystal mounted at a 45 degree angle on the tip of the probe is activated with a 20-MHz range-gated pulsed Doppler, and velocity is determined from Fourier analysis of quadrature data using computer software with a 50-KHz processing range. The device is designed for application to the ascending aorta at the conclusion of a cardiac surgical procedure; and, since mean diameter changes in the ascending aorta have been shown to be very small over a wide range of mean aortic pressures, cardiac output is linearly related to velocity. To test this technique, simultaneous recordings were made from the pulsed Doppler probe and an electromagnetic flow probe in five mongrel dogs over a range of cardiac outputs from 0.5 to 6 l/min for a total of 136 data points. Excellent correlation was found between the electromagnetic flow probe and pulsed Doppler probe data (average r = .98 +/- .006). Average slope between the two was 1.004 and average zero intercept was -0.025 l/min. The catheter is small, stable, has no baseline drift, and continuously and accurately reflects pulsatile changes at high and low flows.

摘要

为解决术后心脏病人心输出量监测的问题,已构建了一种可移除的腔外脉冲多普勒探头,它通过监测升主动脉中的血流速度持续反映心输出量的变化。速度与多普勒频移的频率以及超声束与血管相交的角度有关。一个1平方毫米的压电晶体以45度角安装在探头尖端,由一个20兆赫兹的距离选通脉冲多普勒激活,速度通过使用具有50千赫兹处理范围的计算机软件对正交数据进行傅里叶分析来确定。该装置设计用于在心脏外科手术结束时应用于升主动脉;并且,由于已表明在很宽的平均主动脉压力范围内升主动脉的平均直径变化非常小,心输出量与速度呈线性相关。为测试该技术,在5只杂种狗身上,在0.5至6升/分钟的心输出量范围内,从脉冲多普勒探头和电磁流量探头同时进行记录,共获得136个数据点。在电磁流量探头和脉冲多普勒探头数据之间发现了极好的相关性(平均r = 0.98 ± 0.006)。两者之间的平均斜率为1.004,平均零截距为 -0.025升/分钟。该导管体积小、稳定,无基线漂移,能持续且准确地反映高低流量下的脉动变化。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验