Burghardt R D, Paley D, Specht S C, Herzenberg J E
International Center for Limb Lengthening, Rubin Institute for Advanced Orthopedics, Sinai Hospital of Baltimore, Baltimore, Maryland 21215, USA.
J Bone Joint Surg Br. 2012 Sep;94(9):1241-5. doi: 10.1302/0301-620X.94B9.28672.
Internal lengthening devices in the femur lengthen along the anatomical axis, potentially creating lateral shift of the mechanical axis. We aimed to determine whether femoral lengthening along the anatomical axis has an inadvertent effect on lower limb alignment. Isolated femoral lengthening using the Intramedullary Skeletal Kinetic Distractor was performed in 27 femora in 24 patients (mean age 32 years (16 to 57)). Patients who underwent simultaneous realignment procedures or concurrent tibial lengthening, or who developed mal- or nonunion, were excluded. Pre-operative and six-month post-operative radiographs were used to measure lower limb alignment. The mean lengthening achieved was 4.4 cm (1.5 to 8.0). In 26 of 27 limbs, the mechanical axis shifted laterally by a mean of 1.0 mm/cm of lengthening (0 to 3.5). In one femur that was initially in varus, a 3 mm medial shift occurred during a lengthening of 2.2 cm. In a normally aligned limb, intramedullary lengthening along the anatomical axis of the femur results in a lateral shift of the mechanical axis by approximately 1 mm for each 1 cm of lengthening.
股骨内延长装置沿解剖轴延长,可能导致机械轴向外移位。我们旨在确定沿解剖轴进行股骨延长是否会对下肢对线产生意外影响。使用髓内骨骼动力撑开器对24例患者的27个股骨进行了单纯股骨延长(平均年龄32岁(16至57岁))。排除了同时进行矫正手术或同期胫骨延长,或出现骨不连或骨愈合不良的患者。术前和术后6个月的X线片用于测量下肢对线情况。平均延长长度为4.4厘米(1.5至8.0厘米)。在27条肢体中的26条中,机械轴向外移位,平均每延长1厘米移位1.0毫米(0至3.5毫米)。在1条初始为内翻的股骨中,在延长2.2厘米的过程中出现了3毫米的向内移位。在一条对线正常的肢体中,沿股骨解剖轴进行髓内延长时,每延长1厘米,机械轴向外移位约1毫米。