Moss S M, Marchbanks R J, Burge D M
Institute of Sound and Vibration Research, University of Southampton, England.
Z Kinderchir. 1990 Dec;45 Suppl 1:26-8. doi: 10.1055/s-2008-1042630.
A technique which was originally developed for measuring cochlear fluid pressure, has been under trial for the past two years as a means of assessing ventricular shunt function. The principle of the technique has been reported in previous papers (1, 2, 3) and has been shown to provide a reliable measure of intracranial pressure (ICP) in terms of induced displacement of the tympanic membrane. This study concentrates on the assessment of shunt blockage in spina bifida patients who subsequently underwent shunt revision. The tympanic displacement technique is shown to be of value to this patient group in three respects. Firstly as a research technique to study group-averaged ICP measurements where the ICP needs to be known but invasive measurements cannot be justified. Secondly for assessing shunt dysfunction in individual patients and, finally, in determining the success of shunt revision surgery. The technique was found to be extremely sensitive to relative changes in ICP with shunt revision or any subsequent blockage. The technique is more reliable diagnostically in instances where the patients act as their own controls and a comparison with baseline measurements can be made. Three case reports are illustrated which were selected from those patients tested to emphasise the importance of serial measurements of ICP over period of several days post-revision. The tympanic displacement technique is shown to provide a practical and acceptable method by which this can be undertaken.
一种最初用于测量耳蜗内液体压力的技术,在过去两年中一直在进行试验,作为评估脑室分流功能的一种方法。该技术的原理已在先前的论文(1、2、3)中报道,并且已证明通过鼓膜的诱导位移能够可靠地测量颅内压(ICP)。本研究集中于对脊柱裂患者分流堵塞情况的评估,这些患者随后接受了分流修正手术。鼓膜位移技术在三个方面对该患者群体具有价值。首先,作为一种研究技术,用于研究平均ICP测量值,此时需要知道ICP,但侵入性测量不合理。其次,用于评估个体患者的分流功能障碍,最后,用于确定分流修正手术的成功率。研究发现该技术对分流修正或任何后续堵塞导致的ICP相对变化极其敏感。在患者作为自身对照且可与基线测量值进行比较的情况下,该技术在诊断上更可靠。文中展示了三个病例报告,这些病例选自接受测试的患者,以强调在修正后几天内对ICP进行连续测量的重要性。鼓膜位移技术被证明是一种可行且可接受的进行此项测量的方法。