Gadaleta Matteo, Giorgio Agostino
Dipartimento di Elettrotecnica ed Elettronica, Politecnico di Bari, Via E. Orabona, 4 70125 Bari, Italy.
ISRN Cardiol. 2012;2012:258769. doi: 10.5402/2012/258769. Epub 2012 Aug 26.
This study proposes a method for ventricular late potentials (VLPs) detection using time-frequency representation and wavelet denoising in high-resolution electrocardiography (HRECG). The analysis is performed both with the signal averaged electrocardiography (SAECG) and in real time. A comparison between the temporal and the time-frequency analysis is also reported. In the first analysis the standard parameters QRSd, LAS40, and RMS40 were used; in the second normalized energy in time-frequency domain was calculated. The algorithm was tested adding artificial VLPs to real ECGs.
本研究提出了一种在高分辨率心电图(HRECG)中使用时频表示和小波去噪来检测心室晚电位(VLP)的方法。该分析通过信号平均心电图(SAECG)进行,并且是实时进行的。还报告了时域分析和时频分析之间的比较。在第一次分析中,使用了标准参数QRSd、LAS40和RMS40;在第二次分析中,计算了时频域中的归一化能量。该算法通过向真实心电图中添加人工VLP进行了测试。