Radicioni M, Mezzetti D, Del Vecchio A, Motta M
Neonatal Intensive Care - Paediatric Clinic, S. Maria della Misericordia Hospital of Perugia, Perugia, Italy.
J Matern Fetal Neonatal Med. 2012 Oct;25 Suppl 4:18-21. doi: 10.3109/14767058.2012.714996.
To review the working principles of thromboelastography and evaluate the current knowledge about the possibility of its implementation in the neonatal intensive care unit setting.
Thorough search of the literature in the PubMed database (until May 31, 2012) concerning Thromboelastography (TEG) and/or Thromboelastometry (ROTEM) use in the newborn infant.
Neonatal data are limited to a small number of healthy subjects and virtually absent in the extreme premature infant. Healthy newborns exhibit age dependent accelerated initiation and propagation of coagulation despite prolonged standard plasma coagulation test results, whereas clot firmness and fibrinolysis are similar to adult values. Several neonatal pathological conditions (e.g. sepsis; hypothermia) are linked with substantial changes in the thromboelastographic parameters.
The current knowledge is too limited to express a definitive indication on the reliability of the use of viscoelastic point of care analyzer in the neonatal intensive care unit setting. However, their potential use not only as a diagnostic tool, but also to guide the transfusion therapy requires careful consideration.
回顾血栓弹力图的工作原理,并评估目前关于其在新生儿重症监护病房应用可能性的相关知识。
全面检索PubMed数据库(截至2012年5月31日)中有关新生儿使用血栓弹力图(TEG)和/或血栓弹力测定法(ROTEM)的文献。
新生儿数据仅限于少数健康受试者,在极早产儿中几乎没有相关数据。尽管标准血浆凝血试验结果延长,但健康新生儿的凝血启动和传播呈现出与年龄相关的加速,而凝块硬度和纤维蛋白溶解与成人值相似。几种新生儿病理状况(如败血症;体温过低)与血栓弹力图参数的显著变化有关。
目前的知识非常有限,无法就粘弹性即时检测分析仪在新生儿重症监护病房应用的可靠性给出明确指征。然而,其潜在用途不仅作为诊断工具,还用于指导输血治疗,这需要仔细考虑。