[Influence on measurements of pre-irradiation due to differences in ionization chamber shape or frequency in use].

作者信息

Shimono Tetsunori, Nambu Hidekazu, Matsubara Kosuke, Koshida Kichiro, Gomi Tsutomu

机构信息

Department of Radiology, Hoshigaoka Koseinenkin Hospital.

出版信息

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2012;68(8):986-96. doi: 10.6009/jjrt.2012_jsrt_68.8.986.

Abstract

Ionization chamber measurements in radiation therapy should be repeatedly performed until a stable reading is obtained. Ionization chambers exhibit a response which depends on time elapsed since the previous irradiation. In this study, we investigated the response of a set of two Farmer-style, one Plane parallel, and seven small ionization chambers, which are exposed to 4, 6, 10, and 14 MV. The results show that Farmer-style and Plane parallel ionization chambers settle quickly within 9-20 min. On the other hand, small ionization chambers exhibit settling times of 12-33 min for 6, 10, and 14 MV. It will take longer for a settling time of 4 MV. The settling time showed time dependent irradiation. The first reading was up to 0.76% lower in the Farmer-style and Plane parallel ionization chambers. The small ionization chambers had a 2.60% lower first reading and more gradual response in reaching a stable reading. In this study, individual ionization chambers can vary significantly in their settling behavior. Variation of the responses on ionization chambers were confirmed not only when radiation was not used for a week but also when it was halted for a month. Pre-irradiation of small ionization chambers is clearly warranted for eliminating inadvertent error in the calibration of radiation beams.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验