CO2 microbubble contrast enhancement in x-ray angiography.

机构信息

Department of Radiology, Kansai Medical University, Hirakata, Osaka, Japan.

出版信息

Clin Radiol. 2013 Apr;68(4):346-51. doi: 10.1016/j.crad.2012.08.011. Epub 2012 Sep 13.

Abstract

AIM

To demonstrate that carbon dioxide (CO2) microbubble contrast enhancement depicts blood vessels when used for x-ray examinations.

MATERIALS AND METHODS

Microbubbles were generated by cavitation of physiological saline to which CO2 gas had been added using an ejector-type microbubble generator. The input pressure values for CO2 gas and physiological saline that produced a large quantity of CO2 microbubbles were obtained in a phantom. In an animal study, angiography was performed in three swine using three types of contrast: CO2 microbubbles, conventional CO2 gas, and iodinated contrast medium. For CO2 microbubble contrast enhancement, physiological saline, and CO2 gas were supplied at the input pressures calculated in the phantom experiment. Regions of interest were set in the abdominal aorta, external iliac arteries, and background. The difference in digital values between each artery and the background was calculated.

RESULTS

The input pressures obtained in the phantom experiment were 0.16 MPa for physiological saline and 0.5 MPa for CO2 gas, with physiological saline input volume being 8.1 ml/s. Three interventional radiologists all evaluated the depictions of all arteries as "present" in the CO2 microbubble contrast enhancement, conventional CO2 contrast enhancement, and iodinated contrast enhancement performed in three swine. Digital values for all vessels with microbubble CO2 contrast enhancement were higher than background values.

CONCLUSIONS

In x-ray angiography, blood vessels can be depicted by CO2 microbubble contrast enhancement, in which a large quantity of CO2 microbubbles is generated within blood vessels.

摘要

目的

证明二氧化碳(CO2)微泡造影剂在 X 射线检查中用于描绘血管。

材料和方法

通过向添加了 CO2 气体的生理盐水进行空化来生成微泡,使用喷射式微泡发生器。在体模中获得了产生大量 CO2 微泡的 CO2 气体和生理盐水的输入压力值。在动物研究中,使用三种对比剂(CO2 微泡、常规 CO2 气体和碘对比剂)对三只猪进行了血管造影。对于 CO2 微泡对比增强,以在体模实验中计算出的输入压力供应生理盐水和 CO2 气体。在腹主动脉、外髂动脉和背景中设置感兴趣区域。计算每个动脉与背景之间的数字值差异。

结果

在体模实验中获得的输入压力分别为生理盐水 0.16 MPa 和 CO2 气体 0.5 MPa,生理盐水输入量为 8.1 ml/s。三位介入放射科医生均评估了在三只猪中进行的 CO2 微泡对比增强、常规 CO2 对比增强和碘对比增强中所有动脉的显示情况,均评价为“存在”。所有带有微泡 CO2 对比增强的血管的数字值均高于背景值。

结论

在 X 射线血管造影中,通过 CO2 微泡对比增强可以描绘血管,其中在血管内产生大量 CO2 微泡。

相似文献

1
CO2 microbubble contrast enhancement in x-ray angiography.二氧化碳微泡在 X 射线血管造影中的增强作用。
Clin Radiol. 2013 Apr;68(4):346-51. doi: 10.1016/j.crad.2012.08.011. Epub 2012 Sep 13.
2
Re: CO2 microbubble contrast enhancement in x-ray angiography.关于:X射线血管造影术中的二氧化碳微泡对比增强
Clin Radiol. 2013 Nov;68(11):1179-80. doi: 10.1016/j.crad.2013.06.006. Epub 2013 Jul 26.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验