Department of Cardiology, Aarhus University Hospital, Denmark.
Thromb Res. 2012 Nov;130(5):753-8. doi: 10.1016/j.thromres.2012.08.312. Epub 2012 Sep 17.
Whole blood platelet function testing is widely used to evaluate the effect of antiplatelet agents. Most platelet function tests employ agonists to evaluate drug-specific platelet-activating pathways. Shear-based tests may better reflect physiological conditions in vivo compared with agonist-based tests. The novel PlaCor PRT provides a global platelet reactivity test based on shear-induced platelet aggregation. We evaluated the PlaCor PRT and its agreement with two widely used platelet function tests.
We evaluated platelet function in 85 patients with angiographically verified coronary artery disease on aspirin mono-therapy. Blood sampling was performed one hour after aspirin ingestion. The following platelet function tests were used: PlaCor PRT, multiple electrode aggregometry (Multiplate, agonists: arachidonic acid 1.0 mmol/L and collagen 1.0 μg/mL) and VerifyNow Aspirin.
PlaCor PRT results correlated significantly with multiple electrode aggregometry induced by arachidonic acid (r=-0.38, p<0.001), but not with multiple electrode aggregometry induced by collagen (r=-0.05, p=0.64) or with turbidimetric aggregometry performed with VerifyNow Aspirin (r=-0.13, p=0.23). There was a strong correlation between PlaCor PRT results and platelet count (r=-0.60, p<0.0001), whereas no correlation with haematocrit or von Willebrand factor antigen was found. A relatively large number of instrument errors occurred with the PlaCor PRT device in this initial testing.
PlaCor PRT is a portable, rapid and user-friendly point-of-care test. It shows a modest agreement with comparable tests, and results depend considerably on platelet count. Further validation of the PlaCor PRT is warranted.
全血血小板功能检测广泛用于评估抗血小板药物的效果。大多数血小板功能检测采用激动剂来评估药物特异性的血小板激活途径。与基于激动剂的检测相比,基于剪切的检测可能更能反映体内的生理状况。新型的 PlaCor PRT 提供了一种基于剪切诱导血小板聚集的整体血小板反应性检测。我们评估了 PlaCor PRT 及其与两种广泛使用的血小板功能检测的一致性。
我们评估了 85 例经血管造影证实的冠心病患者在阿司匹林单药治疗下的血小板功能。在服用阿司匹林后 1 小时进行血液采样。使用了以下血小板功能检测:PlaCor PRT、多电极聚集仪(Multiplate,激动剂:花生四烯酸 1.0mmol/L 和胶原 1.0μg/mL)和 VerifyNow 阿司匹林。
PlaCor PRT 结果与多电极聚集仪诱导的花生四烯酸呈显著相关(r=-0.38,p<0.001),但与多电极聚集仪诱导的胶原无相关性(r=-0.05,p=0.64)或与 VerifyNow 阿司匹林进行的比浊聚集试验无相关性(r=-0.13,p=0.23)。PlaCor PRT 结果与血小板计数呈强烈相关(r=-0.60,p<0.0001),而与红细胞压积或血管性血友病因子抗原无关。在最初的测试中,PlaCor PRT 设备出现了相对较多的仪器误差。
PlaCor PRT 是一种便携式、快速且易于使用的即时检测方法。它与可比检测具有一定的一致性,结果在很大程度上取决于血小板计数。需要进一步验证 PlaCor PRT。