Watanabe Hiroyuki, Ito Takeshi, Nakazeko Kazuma, Tachibana Atsuhi, Hashimoto Takeyuki, Shinohara Hiroyuki
Department of Radiology, Showa University Northern Yokohama Hospital, Kanagawa, Japan.
Igaku Butsuri. 2012;31(3):65-74.
In this article, the authors propose an image registration program of portal images and digitally reconstructed radiography (DRR) images used as simulation images for external beam radiation therapy planning. First, the center of the radiation field in a portal image taken using a computed radiograhy cassette is matched to the center of the portal image. Then scale points projected on a DRR image and the portal image are deleted, and the portal image with the radiation field is extracted. Registration of the DRR and portal images is performed using mutual information as the registration criterion. It was found that the absolute displacement misregistrations in two directions (x, y) were 1.2 +/- 0.7mm and 0.5 +/- 0.3 mm, respectively, and rotation disagreement about the z axis 0.3 +/- 0.3 degrees. It was concluded the proposed method was applicable to image registration of portal and DRR images in radiation therapy.
在本文中,作者提出了一种用于外照射放射治疗计划的模拟图像——射野图像与数字重建X线摄影(DRR)图像的图像配准程序。首先,将使用计算机X线摄影暗盒拍摄的射野图像中的辐射野中心与射野图像的中心进行匹配。然后,删除投影在DRR图像和射野图像上的比例尺点,并提取带有辐射野的射野图像。使用互信息作为配准标准来进行DRR图像和射野图像的配准。结果发现,在两个方向(x、y)上的绝对位移配准误差分别为1.2±0.7毫米和0.5±0.3毫米,绕z轴的旋转不一致为0.3±0.3度。得出的结论是,所提出的方法适用于放射治疗中射野图像和DRR图像的图像配准。