Abstract

Distal locking of intramedullary nails (IMNs) is a difficult part of intramedullary nailing (IMN) that could be time-consuming and expose the surgeon, the surgery personnel and the patient to a considerable amount of radiation as fluoroscopy is usually guiding the procedure. Utilization of electromagnetic fields for that purpose offers an attractive alternative. The SURESHOT™ Distal Targeting System (Smith & Nephew, Inc., Memphis, TN, USA) is a novel commercially available radiation-free aiming system that utilizes computerized electromagnetic field tracking technology for the distal locking of IMNs. In order to evaluate the efficacy of the system we conducted the present study. Nineteen patients (six females-thirteen males, mean age 39.5 years, range 17-85 years) with closed diaphyseal fracture of the femur (eight patients) or the tibia (eleven patients) were treated with IMN using the SURESHOT™ Distal Targeting System for the distal interlocking. All targeting attempts were successful at first try and followed by correct positioning of the screws. Mean time for distal locking of tibial IMNs (two screws) was 219sec (range 200-250sec). Mean time for distal locking of femoral IMNs (two screws) was 249 (range 220-330sec). In the current study the SURESHOT™ Distal Targeting System proved to be accurate, fast and easy to learn.

摘要

交锁髓内钉(IMN)的远端锁定是髓内钉的一个难点,过程耗时,透视引导往往会使术者、手术人员和患者暴露在大量辐射下。而使用电磁场来实现这个目标是一个很有吸引力的替代方案。SURESHOT™ 远端瞄准系统(Smith & Nephew, Inc.,田纳西州孟菲斯,美国)是一种新颖的、市售的无辐射瞄准系统,它利用计算机电磁场跟踪技术进行 IMN 的远端锁定。为了评估该系统的效果,我们进行了本研究。19 名患者(6 名女性,13 名男性;平均年龄 39.5 岁,范围 17-85 岁),股骨(8 例)或胫骨(11 例)闭合性骨干骨折,使用 SURESHOT™ 远端瞄准系统进行 IMN 的远端交锁。所有的瞄准尝试都是第一次成功,随后正确定位了螺钉。胫骨 IMN(两个螺钉)的远端锁定时间平均为 219 秒(范围 200-250 秒)。股骨 IMN(两个螺钉)的远端锁定时间平均为 249 秒(范围 220-330 秒)。在本研究中,SURESHOT™ 远端瞄准系统被证明是准确、快速且易于学习的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验