School of Pharmacy, Shenyang Pharmaceutical University, Shenyang 110016, China.
J Pharm Biomed Anal. 2013 Jan;72:261-6. doi: 10.1016/j.jpba.2012.09.002. Epub 2012 Sep 12.
A simple and rapid high performance liquid chromatography electrospray ionization ion-trap tandem mass spectrometry (LC-MS/MS) method has been developed and validated for the quantitative determination of esculentoside A (EsA) in dog plasma using ginsenoside Rg1 as the internal standard (IS). After liquid-liquid extraction (LLE) with n-butanol, the analyte and IS were separated on a Diamonsil C(18) (2.1 mm × 50 mm, 3 μm) column with the mobile phase of methanol-water containing 0.1% acetic acid (70:30, v/v) at a flow rate of 0.2 ml/min. An ion trap mass spectrometer equipped with an electrospray ionization source performed in selected reaction monitoring (SRM) mode was used as the detector. The precursor-product ion transitions were m/z 849.3 M+Na→m/z 805.3 for EsA and m/z 823.3 M+Na→m/z 643.3 for IS. The total chromatographic run time was 5 min. The method was sensitive enough with a lower limit of quantitation (LLOQ) of 5 ng/ml and had a good linearity (r(2)>0.997) over the linear range of 5-500 ng/ml. The mean extraction recovery of EsA from spiked plasma samples was over 75%. The intra- and inter-precisions were no more than 8.8% and accuracies were within the range of -4.6 to 8.7%. All the validated data were within the accepted criteria as stated in the FDA bioanalytical method validation guideline. The developed method was suitable for the quantification of EsA and successfully applied to the pharmacokinetic study of EsA after an oral administration to beagle dogs.
建立并验证了一种简单、快速的高效液相色谱-电喷雾离子阱串联质谱(LC-MS/MS)法,用于狗血浆中竹节参皂苷 A(EsA)的定量分析,以人参皂苷 Rg1 为内标(IS)。采用正丁醇进行液液萃取(LLE)后,用甲醇-水(含 0.1%乙酸,70:30,v/v)作为流动相,在 Diamonsil C(18)(2.1mm×50mm,3μm)柱上进行分离,流速为 0.2ml/min。采用电喷雾离子源的离子阱质谱仪在选择反应监测(SRM)模式下作为检测器。前体-产物离子转换为 m/z 849.3 M+Na→m/z 805.3(用于 EsA)和 m/z 823.3 M+Na→m/z 643.3(用于 IS)。总色谱运行时间为 5min。该方法足够灵敏,定量下限(LLOQ)为 5ng/ml,线性范围为 5-500ng/ml,线性良好(r(2)>0.997)。竹节参皂苷 A 从加标血浆样品中的平均提取回收率超过 75%。内、日间精密度均不超过 8.8%,准确度在-4.6%至 8.7%范围内。所有验证数据均符合 FDA 生物分析方法验证指南中规定的可接受标准。所建立的方法适用于 EsA 的定量分析,并成功应用于口服给予比格犬后 EsA 的药代动力学研究。