Sohrabi K, Basu D, Schudt F, Scholtes M, Balzer B, Hildebrandt O, Koehler U, Gross V
ThoraTech GmbH, Anwenderzentrum Medizintechnik, Gießen.
Pneumologie. 2012 Nov;66(11):669-73. doi: 10.1055/s-0032-1325692. Epub 2012 Sep 26.
Respiratory flow detection with the aim of detecting sleep-related breathing disorders plays a major role in polysomnography. Due to the fact that pneumotachographs are too bulky and not suitable for measurements during sleep, the ThorAKUSTIK system has been developed. By attaching a noise sensor right next to larynx, it determines the respiratory flow in an acoustic way.
The ThorAKUSTIK system as well as a pneumotachograph were applied simultaneously. The correlation between those two methods has been calculated.
We investigated twenty male subjects. All of them were non-smokers.
The ThorAKUSTIK-System showed a highly positive correlation (r = 0.89 to 0.91; p < 0.01) and was able to measure the respiratory flow in a reliable way.
The ThorAKUSTIK-System allows a long-term live monitoring and has the potential to be used in several clinical departments. Larger studies are necessary to verify the application in the clinical routine.