Suppr超能文献

[近端指间关节(PIP)手指假体——我们学到了什么?十年经验]

[Proximal interphalangeal (PIP) finger prosthesis - what have we learnt? Experiences over 10 years].

作者信息

Vogt R, Aerni M, Ampofo C, Schmelzer-Schmied N

机构信息

Kantonsspital Münsterlingen, Klinik für Chirurgie, Münsterlingen, Schweiz.

出版信息

Handchir Mikrochir Plast Chir. 2012 Oct;44(5):293-9. doi: 10.1055/s-0032-1323841. Epub 2012 Oct 1.

Abstract

PURPOSE

In the following article we evaluate the clinical and radiological results after PIP joint replacement to answer the question if the outcome improved over the last years.

PATIENTS AND METHODS

In a retrospective clinical study 53 patients with 66 PIP joint implants (SBI-Avanta, Small Bone Innovations, Inc.) were assessed on the basis of clinical and radiological parameters. We examined the patients' range of motion, their grip strength, the rate of radiological loosening, the development of pain and the patients' satisfaction. 2 groups were included in the study. Patients operated from 2001 to 2007 (group A) were compared with patients operated from 2008 to 2011 (group B). To facilitate a standardized examination, we developed a score providing a comprehensive representation of the outcome after PIP joint replacement.

RESULTS

Group B showed significant improvement in the rate of loosening of the prosthesis (44-4%) in the last follow-up examination. Further improvement was demonstrated in pain reduction, range of motion (77°/67°), the patients' satisfaction and development of swan neck deformity (11%/18%). There was no significant improvement in the patients' grip strength and the rate of revision surgery.

CONCLUSION

Overall, the results after PIP joint replacement are very good. All patients reported a marked improvement in pain reduction. In group B a significant reduction in the rate of loosening of the prosthesis was shown. Moreover, the rate of swan neck deformity, the range of motion, the reduction of pain and the patient's satisfaction improved over the last years. The application of a new PIP-joint-score for a standardized evaluation of the outcome indicated a good correlation with patient satisfaction over time. The experience of the past years enabled improvements in PIP joint replacement. We consider cementing the prosthesis, a large contact area between the prosthesis and the bone and an emphasis on easy intraoperative flexion/extension as factors which can improve the results after PIP joint replacement. However, some problems could not been solved to date. In this respect, prospective clinical and biomechanical studies are necessary.

摘要

目的

在以下文章中,我们评估了近端指间关节置换术后的临床和放射学结果,以回答在过去几年中其结果是否有所改善这一问题。

患者与方法

在一项回顾性临床研究中,基于临床和放射学参数对53例患者的66个近端指间关节植入物(SBI - Avanta,Small Bone Innovations公司)进行了评估。我们检查了患者的活动范围、握力、放射学松动率、疼痛发展情况以及患者满意度。该研究纳入了2组患者。将2001年至2007年接受手术的患者(A组)与2008年至2011年接受手术的患者(B组)进行比较。为便于进行标准化检查,我们制定了一个评分系统,以全面呈现近端指间关节置换术后的结果。

结果

在最后一次随访检查中,B组假体松动率有显著改善(从44%降至4%)。在疼痛减轻、活动范围(77°/67°)、患者满意度以及鹅颈畸形发展情况(11%/18%)方面也有进一步改善。患者的握力和翻修手术率没有显著改善。

结论

总体而言,近端指间关节置换术后的结果非常好。所有患者均报告疼痛明显减轻。B组显示假体松动率显著降低。此外,在过去几年中,鹅颈畸形率、活动范围、疼痛减轻程度以及患者满意度均有所改善。应用新的近端指间关节评分系统对结果进行标准化评估表明,其与患者满意度随时间的变化具有良好的相关性。过去几年的经验使近端指间关节置换术得以改进。我们认为假体的骨水泥固定、假体与骨之间较大的接触面积以及强调术中易于屈伸是可以改善近端指间关节置换术后结果的因素。然而,一些问题至今仍未解决。在这方面,有必要进行前瞻性的临床和生物力学研究。