置于胸部的全植入式静脉通路装置的不透射线情况及偶然发现的机械并发症。

Radio-opacity and incidental identified mechanical complications of totally implantable venous access devices placed in the chest.

作者信息

Surov Alexey, Rusner Carsten, Weigand Karl, John Endres, Spielmann Rolf Peter, Behrmann C

机构信息

Department of Radiology, Martin-Luther-University Halle-Wittenberg, Halle, Germany.

出版信息

Acta Radiol. 2012 Nov 1;53(9):1035-9. doi: 10.1258/ar.2012.110644. Epub 2012 Oct 1.

Abstract

BACKGROUND

Totally implantable venous access devices (TIVAD) may be associated with different complications. Certain mechanical port disorders can easily be diagnosed on chest radiographs if the implanted systems are radiopaque and well visible. There are no reports regarding the visibility of TIVAD on chest X-rays.

PURPOSE

To assess the radio opacity of TIVAD implanted in the chest as well as type and frequency of mechanical complications of ports on chest X-ray images.

MATERIAL AND METHODS

Chest X-rays of 985 patients from the time period 2007-2009 were analyzed retrospectively. In these patients 1190 TIVAD were inserted. All parts of the TIVAD, i.e. port chamber, connection, and port catheter, were checked for their visibility on chest radiographs. An opacity score was used here as follows: ++ well visible; + visible; - partly or completely invisible. Mechanical complications of TIVAD incidentally detected on chest X-ray were also analyzed retrospectively.

RESULTS

Nineteen TIVAD models with diverse configuration and visibility of port chambers, connections, and catheters were identified in our study. Eighty-eight percent of the analyzed port systems were well visible or visible on chest radiographs. Twelve percent of the port chambers and catheters were partly visible or completely invisible. In 9% of the TIVAD, different mechanical complications were diagnosed on chest X-ray images.

CONCLUSION

TIVADs should be evaluated carefully on every chest X-ray. Ideally, they should be radio-opaque and well visible on thoracic X-ray images. Unfortunately, this is not always the case. Therefore, manufacturers of TIVAD should take into consideration to use exclusively radio-opaque materials that allow sufficient visibility of each port component on chest radiographs.

摘要

背景

完全植入式静脉通路装置(TIVAD)可能会引发不同的并发症。如果植入系统具有放射性且在胸部X光片上清晰可见,某些机械端口紊乱很容易被诊断出来。目前尚无关于TIVAD在胸部X光片上可见性的报道。

目的

评估植入胸部的TIVAD的放射性不透光性以及胸部X光图像上端口机械并发症的类型和频率。

材料与方法

回顾性分析了2007年至2009年期间985例患者的胸部X光片。在这些患者中,共插入了1190个TIVAD。检查了TIVAD的所有部件,即端口腔、连接件和端口导管在胸部X光片上的可见性。此处使用的不透光性评分如下:++清晰可见;+可见;-部分或完全不可见。对胸部X光片上偶然发现的TIVAD机械并发症也进行了回顾性分析。

结果

在我们的研究中,识别出了19种具有不同配置以及端口腔、连接件和导管可见性的TIVAD型号。88%的分析端口系统在胸部X光片上清晰可见或可见。12%的端口腔和导管部分可见或完全不可见。在9%的TIVAD中,胸部X光图像上诊断出了不同的机械并发症。

结论

每次胸部X光检查时都应仔细评估TIVAD。理想情况下,它们应具有放射性不透光性且在胸部X光图像上清晰可见。不幸的是,情况并非总是如此。因此,TIVAD制造商应考虑仅使用放射性不透光材料,以使每个端口组件在胸部X光片上具有足够的可见性。