Abstract

The purpose of this study is to validate the capability of in-house independent point dose calculation software to be used as a second check for Helical Tomotherapy treatment plans. The software performed its calculations in homogenous conditions (using the Cheese phantom, which is a cylindrical phantom with radius 15 cm and length 18 cm) using a factor-based algorithm. Fifty patients, who were treated for pelvic (10), prostate (14), lung (10), head & neck (12) and brain (4) cancers, were used. Based on the individual patient kVCT images and the pretreatment MVCT images for each treatment fraction, the corresponding daily patient setup shifts in the IEC-X, IEC-Y, and IEC-Z directions were registered. For each patient, the registered fractional setup shifts were grouped into systematic and random shifts. The average systematic dosimetric variations showed small dose deviation for the different cancer types (1.0%-3.0%) compared to the planned dose. Of the fifty patients, only three had percent differences larger than 5%. The average random dosimetric variations showed relatively small dose deviations (0.2%-1.1%) compared to the planned dose. None of the patients had percent differences larger than 5%. By examining the individual fractions of each patient, it is observed that only in 31 out of 1358 fractions the percent differences exceeded the border of 5%. These results indicate that the overall dosimetric impact from systematic and random variations is small and that the software is a capable platform for independent point dose validation for the Helical Tomotherapy modality.

摘要

本研究旨在验证内部独立点剂量计算软件作为螺旋断层放疗计划的二次检查的能力。该软件在同质条件下(使用奶酪体模,这是一个半径为 15 厘米、长 18 厘米的圆柱形体模)使用基于因子的算法进行计算。共使用了 50 名接受盆腔(10 名)、前列腺(14 名)、肺部(10 名)、头颈部(12 名)和脑部(4 名)癌症治疗的患者。基于每个治疗部分的个体患者千伏计算机断层摄影术(kVCT)图像和预处理兆伏计算机断层摄影术(MVCT)图像,相应的每日患者在 IEC-X、IEC-Y 和 IEC-Z 方向上的设置偏移量被注册。对于每个患者,注册的分次设置偏移量被分为系统和随机偏移量。平均系统剂量学变化显示不同癌症类型的剂量偏差较小(1.0%-3.0%)与计划剂量相比。在这 50 名患者中,只有 3 名患者的百分比差异大于 5%。平均随机剂量学变化显示与计划剂量相比相对较小的剂量偏差(0.2%-1.1%)。没有患者的百分比差异大于 5%。通过检查每个患者的个别部分,可以观察到只有在 1358 个部分中的 31 个部分中,百分比差异超过了 5%的边界。这些结果表明,系统和随机变化的总体剂量学影响较小,该软件是螺旋断层放疗模式独立点剂量验证的可靠平台。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验