McClean B A, Overton T R, Hangartner T N, Rathee S
Department of Applied Sciences in Medicine, University of Alberta, Edmonton, Canada.
Phys Med Biol. 1990 Jan;35(1):11-9. doi: 10.1088/0031-9155/35/1/002.
A special purpose x-ray CT scanner with the capability of scanning objects 75-220 mm in diameter with constant relative geometrical resolution has been developed. The data collection scheme for the scanner uses multiple rotations of a linearly shifted, asymmetric fan beam permitting user-defined variable resolution. Details of hardware and the calibration procedures for the scanner are described and the methods used to measure trabecular bone density (TBD) in the peripheral skeleton are outlined. The standard error of estimate (SEE) of a calibration line of pixel value as a function of K2HPO4 concentration was determined to be 0.07%. The short-term, in vivo precision of the TBD determination, by repeated measurements of a volunteer with repositioning between each measurement, was +/- 0.67% (coefficient of variation (CV] with a 50s scan time and a radiation dose of less than 20 mR per slice.
已开发出一种特殊用途的X射线CT扫描仪,它能够以恒定的相对几何分辨率对直径75至220毫米的物体进行扫描。该扫描仪的数据采集方案采用线性移位非对称扇形束的多次旋转,允许用户定义可变分辨率。描述了扫描仪的硬件细节和校准程序,并概述了用于测量外周骨骼小梁骨密度(TBD)的方法。确定像素值校准线作为K2HPO4浓度函数的估计标准误差(SEE)为0.07%。通过对一名志愿者进行重复测量,每次测量之间重新定位,TBD测定的短期体内精密度为+/-0.67%(变异系数(CV),扫描时间为50秒,每切片辐射剂量小于20 mR)。