Jones L A, Hunter I W
School of Physical and Occupational Therapy, McGill University, Montreal, Canada.
Biol Cybern. 1990;62(4):299-307. doi: 10.1007/BF00201444.
An active servo-system was used to change the stiffness of a manipulandum used in a position-control pursuit-tracking task. The elastic stiffness of the manipulandum connected to the forearm was set by a computer at one of five levels ranging from 0 N/m to 2000 N/m. Subjects were required to track, either by moving their forearm or by generating a force isometrically, a visually presented target whose position changed randomly every second for 100 s. Nonparametric and parametric impulse response functions were calculated between the input (target) and output (force or position) in each tracking condition, and revealed that for all subjects force control was faster than position control when the stiffness of the manipulandum was set at 0 N/m. Subjects were also consistently faster in reaching the target when the stiffness was greater than zero, and were more accurate (steady-state response) when the stiffness of the manipulandum was set at lower rather than higher amplitudes. The parametric impulse response functions revealed that the human operator system was underdamped (0.7) with a natural frequency of approximately 8 rad/s. These findings were interpreted in terms of the responses of the various subsystems (visual, cognitive, contractile, limb mechanics) that comprise the human operator's response.
一个主动伺服系统被用于改变在位置控制追踪任务中使用的操作手柄的刚度。连接到前臂的操作手柄的弹性刚度由计算机设置为从0 N/m到2000 N/m的五个级别之一。要求受试者通过移动前臂或等长产生力来追踪一个视觉呈现的目标,该目标的位置每秒随机变化一次,持续100秒。在每个追踪条件下,计算输入(目标)和输出(力或位置)之间的非参数和参数脉冲响应函数,结果显示,当操作手柄的刚度设置为0 N/m时,对于所有受试者,力控制比位置控制更快。当刚度大于零时,受试者到达目标的速度也始终更快,并且当操作手柄的刚度设置为较低而非较高幅度时,受试者更准确(稳态响应)。参数脉冲响应函数显示,人类操作控制系统的阻尼比为0.7,固有频率约为8 rad/s。这些发现是根据构成人类操作者响应的各个子系统(视觉、认知、收缩、肢体力学)的响应来解释的。