Department of Analytical Chemistry, University of A Coruña, Campus da Zapateira, Rua da Fraga 10, E-15008 A Coruña, Spain.
Talanta. 2012 Nov 15;101:413-9. doi: 10.1016/j.talanta.2012.09.050. Epub 2012 Oct 8.
Blank contamination is a notorious problem in the ultratrace analysis of alkylphenols and bisphenol A. The achievement of low detection limits is complicated due to the high background signals. Furthermore, overestimations and underestimations in the analytical results can occur when blank levels are not stable. Thus, a review of sources of blank contamination in this type of analysis was carried out. Several sources of contamination were identified and useful guidelines are proposed for the determination of these compounds in water samples by liquid chromatography coupled with mass spectrometry. The system contamination was maintained below 0.09 ng (reagent blank) for all compounds and below 0.003 μg L(-1) (procedure blank). The main improvement was obtained by using LC-MS grade solvent in the mobile phase and PTFE syringe filters for the filtration of the sample extracts. Sample handling aspects such as filtration and storage of the water samples were also considered. The filtration of the samples should be avoided because both contamination and adsorption problems were observed when different kinds of filters were assayed. The refrigerated storage of water samples should be limited to 5 days (without addition of methanol) or 8 days (with 5% methanol).
空白污染是烷基酚和双酚 A 的痕量分析中一个众所周知的问题。由于背景信号高,实现低检测限变得复杂。此外,当空白水平不稳定时,分析结果可能会出现高估和低估。因此,对这种分析中空白污染的来源进行了综述。确定了几个污染源,并为通过液相色谱-质谱联用测定水样中的这些化合物提出了有用的指南。对于所有化合物,系统污染均保持在 0.09ng(试剂空白)以下,对于 0.003μg L(-1)(程序空白)以下。通过在流动相中使用 LC-MS 级溶剂和 PTFE 注射器过滤器对样品提取物进行过滤,取得了主要的改进。还考虑了样品处理方面,如样品的过滤和储存。应避免过滤样品,因为当测试不同类型的过滤器时,观察到了污染和吸附问题。如果不添加甲醇,水样的冷藏储存应限制在 5 天内;如果添加 5%甲醇,则应限制在 8 天内。