Electrophoresis. 2013 Feb;34(3):374-82. doi: 10.1002/elps.201200267. Epub 2013 Jan 3.
The use of SPE coupled in-line to CE using electrospray MS detection (in-line SPE-CE-ESI-MS) was investigated for the preconcentration and separation of four UV filters: benzophenone-3, 2,2-dihydroxy-4-methoxybenzophenone, 2,4-dihydroxybenzophenone and 2-phenylbenzimidazole-5-sulphonic acid. First, a CE-ESI-MS method was developed and validated using standard samples, obtaining LODs between 0.06 μg/mL and 0.40 μg/mL. For the in-line SPE-CE-ESI-MS method, three different sorbents were evaluated and compared: Oasis HLB, Oasis MCX, and Oasis MAX. For each sorbent, the main parameters affecting the preconcentration performance, such as sample pH, volume, and composition of the elution plug, and sample injection time were studied. The Oasis MCX sorbent showed the best performance and was used to validate the in-line SPE-CE-ESI-MS methodology. The LODs reached for standard samples were in the range between 0.01 and 0.05 ng/mL with good reproducibility and the developed strategy provided sensitivity enhancement factors between 3400-fold and 34 000-fold. The applicability of the developed methodology was demonstrated by the analysis of UV filters in river water samples.
采用固相萃取(SPE)与毛细管电泳(CE)在线联用,并结合电喷雾质谱检测(in-line SPE-CE-ESI-MS),对四种紫外线吸收剂(二苯甲酮-3、2,2-二羟基-4-甲氧基二苯甲酮、2,4-二羟基二苯甲酮和 2-苯并咪唑-5-磺酸)进行了预浓缩和分离。首先,使用标准样品开发并验证了 CE-ESI-MS 方法,获得的检出限在 0.06 μg/mL 至 0.40 μg/mL 之间。对于在线 SPE-CE-ESI-MS 方法,评估并比较了三种不同的吸附剂:Oasis HLB、Oasis MCX 和 Oasis MAX。对于每种吸附剂,都研究了影响预浓缩性能的主要参数,例如样品 pH 值、洗脱塞体积和组成以及样品注入时间。Oasis MCX 吸附剂表现出最佳性能,并用于验证在线 SPE-CE-ESI-MS 方法学。对于标准样品,达到的检出限在 0.01 至 0.05 ng/mL 范围内,具有良好的重现性,所开发的策略提供了 3400 倍至 34000 倍的灵敏度增强因子。通过对河水样品中紫外线吸收剂的分析,证明了所开发方法的适用性。