铟 - 111二乙三胺五乙酸血流研究以评估外科植入式药物泵输送系统。

Indium-111 DTPA flow study to evaluate surgically implanted drug pump delivery system.

作者信息

Rosenson A S, Ali A, Fordham E W, Penn R D

机构信息

Department of Diagnostic Radiology and Nuclear Medicine, Rush-Presbyterian-St. Luke's Hospital, Chicago, Illinois 60612.

出版信息

Clin Nucl Med. 1990 Mar;15(3):154-6. doi: 10.1097/00003072-199003000-00003.

Abstract

The authors describe a radionuclide technique for evaluating intrathecal and intraventricular infusions by surgically implanted drug pump delivery systems. Sixteen patients underwent flow studies, performed by injecting 500 mu Ci of indium-111 DTPA into the pumps. Early and delayed images were obtained. These studies enabled distinction among functioning pumps, nonfunctioning pumps, and obstructed or occluded catheters. We conclude that indium-111 DTPA flow studies provide an excellent way to assess these drug pump delivery systems.

摘要

作者描述了一种通过手术植入式药物泵输送系统来评估鞘内和脑室内输注的放射性核素技术。16名患者接受了血流研究,方法是向泵内注入500微居里的铟-111二乙三胺五乙酸(DTPA)。获得了早期和延迟图像。这些研究能够区分正常工作的泵、不工作的泵以及阻塞或堵塞的导管。我们得出结论,铟-111 DTPA血流研究为评估这些药物泵输送系统提供了一种极好的方法。