Suppr超能文献

硅油取出联合人工晶状体植入术后的屈光效果。

Refractive outcomes after silicone oil removal and intraocular lens implantation.

机构信息

Department of Ophthalmology, Bascom Palmer Eye Institute, Miami Miller School of Medicine, Miami, Florida 33136, USA.

出版信息

Retina. 2013 Mar;33(3):634-41. doi: 10.1097/IAE.0b013e31826d37e4.

Abstract

PURPOSE

To evaluate the accuracy of intraocular lens (IOL) calculations in eyes undergoing silicone oil removal and IOL implantation using ultrasound biometry and partial coherence interferometry (PCI) (IOLMaster).

METHODS

This was a retrospective consecutive series of eyes that underwent silicone oil removal and IOL implantation from 1999 to 2009. Data were collected on preoperative and intraoperative characteristics that could affect the refractive outcome. The predicted refraction, as measured with ultrasound biometry or PCI, was compared with the postoperative refraction.

RESULTS

Overall, the mean difference between observed and predicted refraction was -0.97 diopters (D). Thirty-five of 64 eyes (55%) were measured by ultrasound biometry, and 29 of 64 eyes (45%) were measured by PCI. The mean (SD) difference between the predicted refraction and the observed spherical equivalent was -1.34 D (2.18 D) when measured by biometry and -0.51 D (2.03 D) when measured by PCI (P = 0.13).

CONCLUSION

Refractive outcomes after silicone oil removal and secondary IOL implantation are fairly accurate, with most ending up slightly myopic. Measurement by PCI may be more accurate than biometry. The IOL power should be selected to yield a target refraction about 0.5 D to 1.25 D more hyperopic than desired, depending on the method used to measure it.

摘要

目的

评估在硅油取出和人工晶状体(IOL)植入术中使用超声生物测量和部分相干干涉测量(PCI)(IOLMaster)进行 IOL 计算的准确性。

方法

这是一项回顾性连续系列研究,纳入了 1999 年至 2009 年期间接受硅油取出和 IOL 植入的患者。收集了可能影响屈光结果的术前和术中特征数据。使用超声生物测量或 PCI 测量的预测屈光度与术后屈光度进行比较。

结果

总体而言,观察到的与预测的屈光度之间的平均差异为-0.97 屈光度(D)。64 只眼中有 35 只(55%)通过超声生物测量进行测量,64 只眼中有 29 只(45%)通过 PCI 进行测量。通过生物测量测量的预测球镜等效值与观察到的球镜等效值之间的平均(SD)差异为-1.34 D(2.18 D),通过 PCI 测量的为-0.51 D(2.03 D)(P = 0.13)。

结论

硅油取出和二次 IOL 植入后的屈光结果相当准确,大多数患者最终为轻度远视。与生物测量相比,PCI 测量可能更准确。IOL 度数的选择应使目标屈光度比期望的远视增加 0.5 D 至 1.25 D,具体取决于用于测量的方法。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验