Abstract

The reaction of protonation of 2-amino-2-(hydroxymethyl)-1,3-propanediol (TRIS) is a suitable one for the calibration of isothermal titration calorimeter (ITC), providing that experimental conditions are appropriately chosen. The conditions and methods for handling experimental data from a nanowatt-ITC are discussed. Also, the binding of Ba(2+) to 18-Crown-6 is successfully used to check the accuracy and precision of the chemical calibration performed with TRIS. This latter reaction has the additional advantage that the data can also be used for a check on the determination of the value of a binding constant. The anomaly of the first injection in ITC is analyzed and, by combining calorimetric and spectroscopic measurements, it is shown that it mainly results from a backlash effect of the syringe plunger rather than from a diffusion effect.

摘要

2-氨基-2-(羟甲基)-1,3-丙二醇(TRIS)的质子化反应是一种适合等温滴定量热法(ITC)校准的反应,只要实验条件得到适当选择。讨论了从纳瓦级 ITC 获得的实验数据的处理条件和方法。此外,成功地使用 Ba(2+)与 18-冠-6 的结合来检查用 TRIS 进行的化学校准的准确性和精密度。后一种反应的额外优点是数据也可用于检查结合常数的确定值。分析了 ITC 中第一次注射的异常现象,并通过结合量热法和光谱测量,表明它主要是由于注射器柱塞的反冲效应而不是扩散效应引起的。