Zacharoulis Dimitris, Lazoura Olga, Sioka Eleni, Potamianos Spyros, Tzovaras George, Nicholls Joanna, Koukoulis George, Habib Nagy
Department of Surgery, University Hospital of Larissa, Larissa, Greece.
J Invest Surg. 2013 Feb;26(1):6-10. doi: 10.3109/08941939.2012.681832. Epub 2012 Dec 28.
The Habib EndoHPB is a bipolar radiofrequency (RF) catheter developed to be introduced across malignant strictures of the bile ducts, so that RF energy can locally ablate the tumor prior to stent placement. This experiment aims to assess the ability of the catheter to coagulate the wall of the common bile duct (CBD) in a porcine model, to establish power requirement and time parameters and correlate them to the depth of thermal injury, and to assess the ease of operation of the device.
The CBD was catheterized using the device in 20 pigs. RF energy was applied to the CBD wall with various generator settings. The pigs were sacrificed 24 hr after the application and the CBD was excised for histological analysis.
The device was easy to handle. Statistically significant correlations between the power, the time of RF application, and the thermal injury depth were found.
The Habib EndoHPB catheter can effectively deliver RF energy intraluminally in the porcine CBD. Clinical studies are warranted in order to define proper settings for safe and efficient use in malignant biliary obstruction.
Habib EndoHPB是一种双极射频(RF)导管,设计用于穿过胆管恶性狭窄部位,以便在放置支架前利用射频能量对肿瘤进行局部消融。本实验旨在评估该导管在猪模型中使胆总管(CBD)壁凝固的能力,确定所需功率和时间参数,并将其与热损伤深度相关联,同时评估该设备的操作简便性。
使用该设备对20头猪的胆总管进行插管。采用不同的发生器设置将射频能量施加于胆总管壁。在施加能量24小时后处死猪,切除胆总管进行组织学分析。
该设备易于操作。发现功率、射频施加时间与热损伤深度之间存在统计学显著相关性。
Habib EndoHPB导管能够在猪的胆总管腔内有效传递射频能量。有必要开展临床研究,以确定在恶性胆道梗阻中安全有效使用的合适设置。