J Chromatogr B Analyt Technol Biomed Life Sci. 2013 Jan 15;913-914:138-46. doi: 10.1016/j.jchromb.2012.12.008. Epub 2012 Dec 20.
For the first time, combination of electromembrane extraction (EME) and dispersive liquid-liquid microextraction (DLLME) followed by gas chromatography-flame ionization detection (GC/FID) was developed for determination of tricyclic antidepressants (TCAs) in untreated human plasma and urine samples. Response surface methodology (RSM) was used for optimization of experimental parameters, so that extraction time of 14min, voltage of 240V, donor phase of 64mM HCl and acceptor phase of 100mM HCl were obtained as optimal extraction conditions. Matrix effect and carry-over were investigated in this work. The results indicated matrix effect for urine and plasma samples in comparison with neat solutions, so match matrix method was used for drawing working calibration curves. However, no carry-over was appeared at the retention time of investigated TCAs (S/N<3). With application of optimized values, good linearity in the range of 2-500μg L(-1) was obtained for TCAs with the correlation of determination values (r(2)) above 0.9968. The limits of detection (S/N=3) for TCAs were found 0.25, 3.0, and 15μg L(-1) in water, urine, and plasma, respectively. The preconcentration factors of TCAs in water, urine, and plasma were from 383 to 1065. The intra- and inter-assay precisions (%) were in the ranges 6.4-11.8% and 6.2-10.8%, respectively, and the intra- and inter-assay accuracies were >86.5%. The results showed that EME-DLLME-GC/FID is a promising combination for analysis of TCAs present at low concentrations in biological matrices.
首次将电膜萃取(EME)与分散液液微萃取(DLLME)相结合,结合气相色谱-火焰离子化检测(GC/FID),用于未经处理的人血浆和尿液样品中三环抗抑郁药(TCAs)的测定。响应面法(RSM)用于优化实验参数,使萃取时间为 14min、电压为 240V、供体相为 64mM HCl 和受体相为 100mM HCl,为最佳萃取条件。本工作考察了基质效应和夹带。结果表明,与纯溶液相比,尿液和血浆样品的基质效应,因此采用匹配基质法绘制工作校准曲线。然而,在所研究的 TCAs 的保留时间(S/N<3)处未出现夹带。采用优化值,在 2-500μg L(-1)范围内,TCAs 具有良好的线性关系,相关系数(r(2))大于 0.9968。TCAs 在水、尿液和血浆中的检测限(S/N=3)分别为 0.25、3.0 和 15μg L(-1)。TCAs 在水、尿液和血浆中的浓缩因子分别为 383-1065。内标和日间精密度(%)分别在 6.4-11.8%和 6.2-10.8%范围内,内标和日间准确度均>86.5%。结果表明,EME-DLLME-GC/FID 是一种很有前途的组合,可用于分析生物基质中低浓度的 TCAs。