A case of a C-stem fracture at the head-neck junction and a review of the literature.

作者信息

Morley David, Starks Ian, Lim Justin

机构信息

City General Hospital, University Hospital of North Staffordshire, Newcastle Road, Stoke-on-Trent ST4 6QG, UK.

出版信息

Case Rep Orthop. 2012;2012:158604. doi: 10.1155/2012/158604. Epub 2012 Dec 9.

Abstract

We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use of large femoral heads with neck adaptors and narrow tapers should be used with caution, especially in heavy, active patients.