Department of Clinical Nutrition and Department of Nursing, Penn State Milton S. Hershey Medical Center, Hershey, Pennsylvania 17033, USA.
JPEN J Parenter Enteral Nutr. 2013 Sep;37(5):652-8. doi: 10.1177/0148607112473340. Epub 2013 Jan 18.
Handheld indirect calorimetry has the potential to allow simple and inexpensive measurement of resting metabolic rate in spontaneously breathing people. However, validation work on these devices is contradictory. The purpose of the current study was to determine the bias and level of agreement of oxygen consumption and resting metabolic rate as measured by a handheld indirect calorimeter against a standard open-circuit indirect calorimetry cart.
One hundred community-living, spontaneously breathing, ambulatory nonobese and obese adults were studied in single sessions by a single investigator. Sequential measurements were undertaken using the handheld indirect calorimeter and the standard metabolic cart. Measurement sequence was varied randomly.
The mean value for oxygen consumption and metabolic rate of the 2 devices was not significantly different. However, agreement between the 2 devices was only 43% in nonobese and obese participants, and there was proportional and fixed bias, with the handheld calorimeter tending to produce a higher value for oxygen consumption and resting metabolic rate. Limits of agreement for resting metabolic rate between the 2 calorimeters were -240 to +300 kcal/d.
Measurements of resting metabolic rate by the handheld indirect calorimeter tested in this study are not equivalent to measurements by standard indirect calorimetry.
手持式间接测热仪具有在自主呼吸人群中简单且经济地测量静息代谢率的潜力。然而,这些设备的验证工作存在矛盾。本研究的目的是确定手持式间接测热仪与标准开路间接测热仪测量的耗氧量和静息代谢率的偏差和一致性水平。
在单次试验中,由一名研究人员对 100 名居住在社区、自主呼吸、活动的非肥胖和肥胖成年人进行了研究。使用手持式间接测热仪和标准代谢箱进行了连续测量。测量顺序随机变化。
2 种设备的耗氧量和代谢率平均值没有显著差异。然而,在非肥胖和肥胖参与者中,2 种设备的一致性仅为 43%,且存在比例和固定偏差,手持式测热仪倾向于产生更高的耗氧量和静息代谢率。2 种测热仪之间静息代谢率的容许限为-240 至+300 kcal/d。
本研究中测试的手持式间接测热仪测量的静息代谢率与标准间接测热法测量的结果不一致。