Suppr超能文献

利用心脏机电模型计算主动脉内球囊反搏过程中心脏能量消耗的定量分析。

Computational quantification of the cardiac energy consumption during intra-aortic balloon pumping using a cardiac electromechanics model.

机构信息

Department of Medical IT Convergence Engineering, Kumoh National Institute of Technology, Gumi, Korea.

出版信息

J Korean Med Sci. 2013 Jan;28(1):93-9. doi: 10.3346/jkms.2013.28.1.93. Epub 2013 Jan 8.

Abstract

To quantify the reduction in workload during intra-aortic balloon pump (IABP) therapy, indirect parameters are used, such as the mean arterial pressure during diastole, product of heart rate and peak systolic pressure, and pressure-volume area. Therefore, we investigated the cardiac energy consumption during IABP therapy using a cardiac electromechanics model. We incorporated an IABP function into a previously developed electromechanical model of the ventricle with a lumped model of the circulatory system and investigated the cardiac energy consumption at different IABP inflation volumes. When the IABP was used at inflation level 5, the cardiac output and stroke volume increased 11%, the ejection fraction increased 21%, the stroke work decreased 1%, the mean arterial pressure increased 10%, and the ATP consumption decreased 12%. These results show that although the ATP consumption is decreased significantly, stroke work is decreased only slightly, which indicates that the IABP helps the failed ventricle to pump blood efficiently.

摘要

为了量化主动脉内球囊反搏 (IABP) 治疗期间的工作量减少,使用间接参数,如舒张期平均动脉压、心率和收缩压峰值的乘积以及压力-容积面积。因此,我们使用心脏机电模型研究了 IABP 治疗期间的心脏能量消耗。我们将 IABP 功能纳入先前开发的心室机电模型中,该模型带有循环系统的集中模型,并研究了不同 IABP 充气量下的心脏能量消耗。当 IABP 在充气水平 5 下使用时,心输出量和每搏量增加 11%,射血分数增加 21%,每搏功减少 1%,平均动脉压增加 10%,ATP 消耗减少 12%。这些结果表明,尽管 ATP 消耗显著减少,但每搏功仅略有减少,这表明 IABP 有助于衰竭的心室有效泵血。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d3dd/3546111/1d26b7a98cf8/jkms-28-93-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验