Tay S, Weinberg L, Peyton P, Story D, Briedis J
Department of Anaesthesia, Austin Hospital, Melbourne, Victoria, Australia.
Anaesth Intensive Care. 2013 Jan;41(1):95-101. doi: 10.1177/0310057X1304100116.
Emerging technologies that reduce the economic and environmental costs of anaesthesia have had limited assessment. We hypothesised that automated control of end-tidal gases, a new feature in anaesthesia machines, will consistently reduce volatile agent consumption cost and greenhouse gas emissions. As part of the planned replacement of anaesthesia machines in a tertiary hospital, we performed a prospective before and after study comparing the cost and greenhouse gas emissions of isoflurane, sevoflurane and desflurane when using manual versus automated control of end-tidal gases. We analysed 3675 general anaesthesia cases with inhalational agents: 1865 using manual control and 1810 using automated control. Volatile agent cost was $18.87/hour using manual control and $13.82/hour using automated control: mean decrease $5.05/hour (95% confidence interval: $0.88-9.22/hour, P=0.0243). The 100-year global warming potential decreased from 23.2 kg/hour of carbon dioxide equivalents to 13.0 kg/hour: mean decrease 10.2 kg/hour (95% confidence interval: 2.7-17.7 kg/hour, P=0.0179). Automated control reduced costs by 27%. Greenhouse gas emissions decreased by 44%, a greater than expected decrease facilitated by a proportional reduction in desflurane use. Automated control of end-tidal gases increases participation in low flow anaesthesia with economic and environmental benefits.
能够降低麻醉经济成本和环境成本的新兴技术尚未得到充分评估。我们推测,麻醉机的一项新功能——呼气末气体自动控制,将持续降低挥发性麻醉剂的消耗成本和温室气体排放。作为一家三级医院计划更换麻醉机的一部分,我们进行了一项前瞻性前后对照研究,比较了在使用呼气末气体手动控制与自动控制时,异氟烷、七氟烷和地氟烷的成本及温室气体排放情况。我们分析了3675例使用吸入性麻醉剂的全身麻醉病例:1865例使用手动控制,1810例使用自动控制。使用手动控制时挥发性麻醉剂成本为每小时18.87美元,使用自动控制时为每小时13.82美元:平均每小时降低5.05美元(95%置信区间:0.88 - 9.22美元/小时,P = 0.0243)。100年全球变暖潜能值从每小时23.2千克二氧化碳当量降至13.0千克:平均每小时降低10.2千克(95%置信区间:2.7 - 17.7千克/小时,P = 0.0179)。自动控制使成本降低了27%。温室气体排放减少了44%,地氟烷使用量的成比例减少促成了大于预期的降幅。呼气末气体自动控制增加了对低流量麻醉技术的采用,带来了经济和环境效益。