Using dual-antenna nanosecond pulse near-field sensing technology for non-contact and continuous blood pressure measurement.

作者信息

Lin Hong-Dun, Lee Yen-Shien, Chuang Bor-Nian

机构信息

Center for Measurement Standards, Industrial Technology Research Institute, No. 321, Sec. 2, Kaung Fu Rd., Hsinchu 30011, Taiwan, R.O.C.

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2012;2012:219-22. doi: 10.1109/EMBC.2012.6345909.

Abstract

Long-term and continuous non-invasive blood pressure monitoring has shown that it is the most important to clinical diagnosis of cardiovascular diseases and personal home health care. Currently, there are many preferable non-invasive methods, including optical sensor, pressure-sensitive transducers, and applanation tonometry, to get insight of blood pressure and flow signal detection. However, the operation of traditional monitors is cuff accessories needed, and also the sensing probes needed to exert pressure to the user directly. The measurement procedure is limited by long-term, continuous measurement and also easy to cause discomfort. To improve these issues, the non-pressurized and non-invasive measuring method will become an important innovation improvement. In this paper, the novel nanosecond pulse near-field sensing (NPNS) based screening technology with dual-antenna, which includes radio frequency (RF) pulse transmission and two combined flat antenna connected to transceiver of miniature radar, is proposed to derive relative blood pressure parameters from measured blood flow activity (Pulse Wave Velocity, PWV). A dedicated analysis software is also provided to calculate cardiovascular parameters, including PWV, average systolic time, diastolic and systolic pressure, for clinical and homecare applications. To evaluate the performance, the proposed method was applied on blood pressure measurement at the body site of limbs (brachial and leg). As a result, it shows error of DBP and SBP is 5.18±1.61 and 4.09 ± 1.69 mmHg in average compared with the measurement result from commercial product, and performs the capability of continuous long-term monitoring in real-time.

摘要

长期连续的无创血压监测已表明,它对心血管疾病的临床诊断和个人家庭健康护理最为重要。目前,有许多较好的无创方法,包括光学传感器、压敏换能器和平板眼压测量法,以深入了解血压和血流信号检测。然而,传统监测器的操作需要袖带配件,并且传感探头需要直接向用户施加压力。测量过程受到长期连续测量的限制,并且容易引起不适。为改善这些问题,非加压无创测量方法将成为一项重要的创新改进。本文提出了一种基于双天线的新型纳秒脉冲近场传感(NPNS)筛选技术,该技术包括射频(RF)脉冲传输以及连接到微型雷达收发器的两个组合平板天线,以从测量的血流活动(脉搏波速度,PWV)中导出相对血压参数。还提供了一个专用分析软件来计算心血管参数,包括PWV、平均收缩时间、舒张压和收缩压,用于临床和家庭护理应用。为评估性能,将所提出的方法应用于肢体(肱部和腿部)身体部位的血压测量。结果表明,与商业产品的测量结果相比,其舒张压和收缩压的平均误差分别为5.18±1.61和4.09±1.69 mmHg,并且具有实时连续长期监测的能力。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验