School of Life Science and BioEngineering, Beijing University of Technology, Beijing, China.
Artif Organs. 2013 Feb;37(2):157-65. doi: 10.1111/j.1525-1594.2012.01579.x.
Intra-aorta pump is a novel rotary ventricular assist device. Because of the special structure and connection with the native heart, the hemodynamic effect of support mode of this pump on the cardiovascular system is not clear. In this work, three support modes, including "constant speed" mode, "co-pulse" mode, and "counter-pulse" mode, have been designed for the intra-aorta pump to evaluate the hemodynamic effect of different support modes on the cardiovascular system. Simulation results demonstrate that that both "co-pulse" mode and "counter-pulse" mode can achieve better unloading performance than "constant speed" mode. The intra-aorta pump controlled by "co-pulse" mode is beneficial for improving coronary flow. Moreover, the external work, which is defined as the product of left ventricular pressure and cardiac output, under "co-pulse" mode is the minimum of the three support modes (0.783 w vs. 0.615 w vs. 0.702 w). The pulsatility ratio, defined as the ratio of the peak-to-peak value of arterial pressure (AP) to the mean arterial pressure value, under "co-pulse" mode is the maximum of the three modes (24% vs. 32.8% vs. 23.7%). The equivalent afterload value, which is the ratio of pulsatile pressure at the pump inflow and pulsatile pump flow, is larger than other support modes (0.596 mm Hg.s/mL vs. 0.9704 mm Hg.s/mL vs. 0.55 mm Hg.s/mL). In brief, the intra-aorta pump under "co-pulse" mode support is beneficial for improving myocardial perfusion and restoring pulsatility of AP, while "counter-pulse" mode is beneficial to the perfusion of vital organs.
主动脉内泵是一种新型的旋转心室辅助装置。由于其特殊的结构和与原生心脏的连接,这种泵的支持模式对心血管系统的血流动力学影响尚不清楚。在这项工作中,设计了三种支持模式,包括“恒速”模式、“合拍”模式和“对拍”模式,用于评估主动脉内泵的不同支持模式对心血管系统的血流动力学影响。模拟结果表明,“合拍”模式和“对拍”模式都能比“恒速”模式实现更好的卸载性能。由“合拍”模式控制的主动脉内泵有利于改善冠状动脉血流。此外,“合拍”模式下的外部功(定义为左心室压力和心输出量的乘积)是三种支持模式中最小的(0.783w 比 0.615w 比 0.702w)。“合拍”模式下的脉动比(定义为动脉压(AP)峰值与平均动脉压值的比值)是三种模式中最大的(24%比 32.8%比 23.7%)。泵入压和脉动泵流量的脉动比(定义为脉动压力与等效后负荷值的比值)大于其他支持模式(0.596mmHg.s/mL 比 0.9704mmHg.s/mL 比 0.55mmHg.s/mL)。总之,“合拍”模式支持的主动脉内泵有利于改善心肌灌注和恢复 AP 的脉动性,而“对拍”模式有利于重要器官的灌注。