Department of Cardiovascular Surgery, Centre Hospitalier Universitaire Vaudois, Lausanne, Switzerland.
Eur J Cardiothorac Surg. 2013 Aug;44(2):370-4; discussion 374. doi: 10.1093/ejcts/ezs710. Epub 2013 Feb 5.
Residual mitral regurgitation after valve repair worsens patients' clinical outcome. Postimplant adjustable mitral rings potentially address this issue, allowing the reshaping of the annulus on the beating heart under echocardiography control. We developed an original mitral ring allowing valve geometry remodelling after the implantation and designed an animal study to assess device effectiveness in correcting residual mitral regurgitation.
The device consists of two concentric rings: one internal and flexible, sutured to the mitral annulus and a second external and rigid. A third conic element slides between the two rings, modifying the shape of the flexible ring. This sliding element is remotely activated with a rotating tool. Animal model: in adult swine, under cardio pulmonary bypass and cardiac arrest, we shortened the primary chordae of P2 segment to reproduce Type III regurgitation and implanted the active ring. We used intracardiac ultrasound to assess mitral regurgitation and the efficacy of the active ring to correct it.
Severe mitral regurgitation (3+ and 4+) was induced in eight animals, 54 ± 6 kg in weight. Vena contracta width decreased from 0.8 ± 0.2 to 0.1 cm; proximal isovelocity surface area radius decreased from 0.8 ± 0.2 to 0.1 cm and effective regurgitant orifice area decreased from 0.50 ± 0.1 to 0.1 ± 0.1 cm(2). Six animals had a reversal of systolic pulmonary flow that normalized following the activation of the device. All corrections were reversible.
Postimplant adjustable mitral ring corrects severe mitral regurgitation through the reversible modification of the annulus geometry on the beating heart. It addresses the frequent and morbid issue of recurrent mitral valve regurgitation.
瓣环成形术后残余二尖瓣反流会使患者的临床预后恶化。植入后可调节二尖瓣环可解决这一问题,它允许在超声心动图控制下,在跳动的心脏上重塑瓣环。我们开发了一种原始的二尖瓣环,允许在植入后重塑瓣环的几何形状,并设计了一项动物研究来评估该装置在纠正残余二尖瓣反流方面的有效性。
该装置由两个同心环组成:一个内部的、柔性的,缝合在二尖瓣环上,另一个外部的、刚性的。第三个锥形元件在两个环之间滑动,改变柔性环的形状。这个滑动元件可以通过一个旋转工具远程激活。动物模型:在成年猪中,在心肺旁路和心脏骤停下,我们缩短了 P2 段的初级腱索,以模拟 III 型反流,并植入了主动环。我们使用心内超声来评估二尖瓣反流和主动环纠正反流的效果。
在 8 只体重为 54±6kg 的动物中,诱导了严重的二尖瓣反流(3+和 4+)。有效反流口面积从 0.50±0.1cm²降至 0.1±0.1cm²。6 只动物的收缩期肺动脉血流发生逆转,在激活装置后恢复正常。所有的纠正都是可逆的。
植入后可调节的二尖瓣环通过在跳动的心脏上可逆地改变瓣环的几何形状来纠正严重的二尖瓣反流。它解决了二尖瓣反流复发这一常见且病态的问题。