Clinical evaluation of a novel 12-hole irrigated tip catheter ablation system for the treatment of typical atrial flutter-results from the Duo FLAIR clinical study.

作者信息

Lakkireddy Dhanunjaya, Sangrigoli Robert, Sloan Stephen, Wiseman Martin, Weiss Raul, Molin Frank, Kanaan Nabil, Awasthi Yashasvi, Mahapatra Srijoy

机构信息

Center for Excellence in Atrial Fibrillation and EP Research-KU Cardiovascular Research Institute, Bloch Heart Rhythm Center, Mid America Cardiology, University of Kansas Hospital and Medical Center, Kansas City, KS 66160, USA.

出版信息

J Interv Card Electrophysiol. 2013 Jun;37(1):97-103. doi: 10.1007/s10840-012-9765-3. Epub 2013 Feb 9.

Abstract

INTRODUCTION

Higher current density at the proximal end of a six-hole catheter is not irrigated and thus could lead to uneven heating and possibly coagulum and charring. This study tested a novel 12-hole irrigated catheter with six additional holes at the proximal end that provides more uniform cooling during atrial flutter (AFL) ablation.

METHODS

A total of 188 patients (28 females, 66 ± 11 years) were treated for typical AFL at 22 sites in the USA and Canada using Cool Path Duo at 50 W and 45 °C at an irrigation rate of 13 ml/min. The results were compared to historical data from an AFL study which used a six-hole catheter with similar design.

RESULTS

A total of 2,725 RF lesions were applied in 188 patients with a mean RF duration of 18.2 ± 11.7 min per procedure. Procedural success was achieved in 96.3 % (181 of 188) acutely and 98.3 % (173 of 176) patients at 3 months. Steam pops occurred in 0.6 % of lesions (15 of 2,725) without coagulum or charring… Compared to the six-hole irrigated tip catheter, the Cool Path Duo catheter delivered more power (33.8 ± 5.9 vs. 29.7 ± 5.2 W, p < 0.0001) at a lower average tip temperature (34.7 ± 1.5 vs. 37.2 ± 2.1 °C, p < 0.0001) with no statistical differences in either 3-month AFL recurrence or complications.

CONCLUSION

Cool Path Duo™ irrigated tip catheter is safe and effective in treating typical AFL. When compared to a six-hole irrigated tip catheter, the Cool Path Duo catheter delivers more power at a lower temperature.