主动脉瓣流体动力学:关于无瓦尔萨尔瓦窦的思考。

Aortic valve hydrodynamics: considerations on the absence of sinuses of Valsalva.

作者信息

Bottio Tomaso, Buratto Edward, Dal Lin Carlo, Lika Alban, Tarzia Vincenzo, Rizzoli Giulio, Gerosa Gino

机构信息

Cardiovascular Institute, University of Padova, Padova, Italy.

出版信息

J Heart Valve Dis. 2012 Nov;21(6):718-23.

Abstract

BACKGROUND AND AIM OF THE STUDY

The study aim was to compare the hydrodynamics of the Carpentier-Edwards Magna 21 (CEM) and St. Jude Medical Biocor-Epic-Supra 21 (SJME) valves at increasing stroke volume and pulse rate in two different aortic conduits, namely straight and with sinuses of Valsalva present.

METHODS

Both valve types were tested in the aortic chamber of the Sheffield pulse duplicator, at rates of 70, 80 and 90 beats/min, and stroke volumes of 50 and 60 ml. The systolic and diastolic performances were each recorded. The leaflet coaptation time, ventricle isovolumetric time and maximum instantaneous flow rate were also recorded.

RESULTS

Regardless of the aortic conduit, CEM valves showed a significantly lower gradient than SJME valves (p < 0.05), and a significantly larger effective orifice area (EOA) (p < 0.05); the latter parameter was unaffected for both valves, at an increasing pulse rate (p > 0.05). The maximum transvalvular flow velocity was significantly higher in the straight conduit for both valves (p < 0.05). With regards to diastole, the SJME valve showed the lowest regurgitant volume (p < 0.05). The leaflet coaptation time was significantly shorter for the SJME valve than for the CEM valve (p < 0.05), but when tested in a straight conduit it was shortened significantly for both valves (p < 0.05).

CONCLUSION

An absence of the sinuses of Valsalva may modify the diastolic and systolic behaviors of the tissue valve leaflets by reducing the time required for leaflet coaptation, and increasing the valve closing volume and maximum transvalvular flow velocity. It is speculated that these hydrodynamic changes may increase the working stress on the valve tissue, leading to possible premature structural valve deterioration.

摘要

研究背景与目的

本研究旨在比较Carpentier-Edwards Magna 21(CEM)瓣膜和圣犹达医疗Biocor-Epic-Supra 21(SJME)瓣膜在两种不同主动脉管道(即直管和存在主动脉窦)中,随着每搏量和脉率增加时的流体动力学情况。

方法

两种瓣膜类型均在谢菲尔德脉搏模拟器的主动脉腔中进行测试,心率分别为70、80和90次/分钟,每搏量分别为50和60毫升。记录收缩期和舒张期的性能。还记录了瓣叶贴合时间、心室等容时间和最大瞬时流速。

结果

无论主动脉管道如何,CEM瓣膜的梯度均显著低于SJME瓣膜(p < 0.05),有效瓣口面积(EOA)显著更大(p < 0.05);随着脉率增加,后一参数对两种瓣膜均无影响(p > 0.05)。两种瓣膜在直管中的最大跨瓣流速均显著更高(p < 0.05)。在舒张期方面,SJME瓣膜的反流容积最低(p < 0.05)。SJME瓣膜的瓣叶贴合时间显著短于CEM瓣膜(p < 0.05),但在直管中测试时,两种瓣膜的瓣叶贴合时间均显著缩短(p < 0.05)。

结论

主动脉窦的缺失可能通过减少瓣叶贴合所需时间、增加瓣膜关闭容积和最大跨瓣流速来改变组织瓣膜瓣叶的舒张期和收缩期行为。据推测,这些流体动力学变化可能会增加瓣膜组织的工作应力,导致瓣膜结构可能过早退化。