Suppr超能文献

在荧光透视系统自动剂量控制的验收测试期间,我们应该在哪里测量空气比释动能率?

Where should we measure the entrance air kerma rate during acceptance testing of the automatic dose control of a fluoroscopic system?

作者信息

Fukuda Atsushi, Miyati Tosiaki, Matsubara Kosuke

机构信息

Department of Radiology, Shiga Medical Center for Children, 5-7-30, Moriyama, Moriyama city, Shiga 524-0022, Japan.

出版信息

Radiol Phys Technol. 2013 Jul;6(2):313-6. doi: 10.1007/s12194-013-0202-9. Epub 2013 Feb 15.

Abstract

In Japan, the entrance air kerma rate (EAKR) to a patient cannot exceed 50 mGy/min in conventional fluoroscopy. However, it is unclear where the EAKR should be measured. We obtained the tube potential and tube current as a function of polymethylmethacrylate (PMMA) thickness, and the EAKR at the interventional reference point (IRP) was measured from the trajectory. The EAKR at the point established by the U.S. Food and Drug Administration (FDA) was calculated from EAKR at the IRP. The EAKR at the IRP exceeded the limit at a PMMA thickness of 22-28 cm. However, the EAKR did not exceed the limit at the FDA point. If the EAKR to a patient is being verified to meet the recent Japanese ruling, the EAKR should be measured at the FDA point, and if the EAKR is being evaluated for determination of the skin dose, it should be monitored at the IRP.

摘要

在日本,传统透视检查中患者的入射空气比释动能率(EAKR)不得超过50毫戈瑞/分钟。然而,尚不清楚应在何处测量EAKR。我们获取了管电压和管电流作为聚甲基丙烯酸甲酯(PMMA)厚度的函数,并从轨迹测量了介入参考点(IRP)处的EAKR。根据IRP处的EAKR计算了美国食品药品监督管理局(FDA)确定的点处的EAKR。当PMMA厚度为22 - 28厘米时,IRP处的EAKR超过了限值。然而,FDA点处的EAKR未超过限值。如果要验证患者的EAKR是否符合日本最新规定,应在FDA点测量EAKR;如果是为了确定皮肤剂量而评估EAKR,则应在IRP处进行监测。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验