Determination of 17 β-Estradiol in Rabbit Plasma by Gas Chromatography with Flame Ionization Detection.

作者信息

Yilmaz B, Kadioglu Y

机构信息

Department of Analytical Chemistry, Faculty of Pharmacy, Ataturk University, 25240, Erzurum, Turkey.

出版信息

Indian J Pharm Sci. 2012 May;74(3):248-53. doi: 10.4103/0250-474X.106067.

Abstract

This article describes gas chromatography-flame ionization detection method for determination of 17 β-estradiol in rabbit plasma. 17 β-estradiol and internal standard progesterone were extracted from plasma using liquid-liquid extraction method. Linearity was found between 0.25 and 20 μg/ml (r(2)=0.994) for plasma samples. Intra-day and inter-day precision, expressed as the relative standard deviation were less than 5.5%, and accuracy (relative error) was less than 3.5%. The mean recovery of 17 β-estradiol samples was 94.4%. The limits of detection and quantification of method for plasma samples were 0.10 μg/ml and 0.15 μg/ml, respectively. Also, clinically used other 10 drugs were investigated to check for potential interferences and the method was successfully applied to the determination of 17 β-estradiol in New Zealand white rabbits.