García Sáez Diana, Mohite Prashant, Zych Bartlomeij, Sabashnikov Anton, Simon André R
Department of Cardiothoracic Transplantation and Mechanical Support, Royal Brompton and Harefield NHS Trust, Harefield Hospital, Harefield, Middlesex, UK.
Interact Cardiovasc Thorac Surg. 2013 Jun;16(6):888-9. doi: 10.1093/icvts/ivt066. Epub 2013 Feb 26.
Despite technological advances in a newer generation of ventricular assist devices (VAD), complications, such as pump thromboses, remain a significant cause of morbidity and indeed mortality in these patients. We present the case of a 34-year old patient who underwent HeartAssist 5 (HA5) implantation as a bridge to cardiac transplant. After an initial uneventful recovery, he developed a pump thrombosis that was refractory to medical treatment. We present the surgical technique used to exchange the HA5 with a HeartWare (HVAD), leaving the old inflow-sewing ring in situ.