艾可乐托比 709 M 散光矫正型人工晶状体的有效性和稳定性。

Efficacy and stability of AT TORBI 709 M toric IOL.

机构信息

Ophthalmology Department, Hospital Universitario Donostia, San Sebastian, Spain.

出版信息

J Refract Surg. 2013 Mar;29(3):194-9. doi: 10.3928/1081597X-20130129-02.

Abstract

PURPOSE

To describe the intraocular rotational stability and refractive results of the toric AT TORBI 709 M (former AT.Comfort 464 TLC) intraocular lens (IOL) (Carl Zeiss Meditec AG, Jena, Germany).

METHODS

Forty-eight eyes of 32 consecutive patients with cataract and greater than 1 diopter of corneal astigmatism were included. Implantation of the toric IOL was performed after phacoemulsification. Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), residual refractive sphere, residual keratometric and refractive cylinders, and toric IOL axis were measured before and 1, 3, and 6 months after surgery.

RESULTS

Mean logMAR UDVA and CDVA improved significantly after surgery (P < .05). UDVA was 20/40 or better in 88.1% of eyes and 20/25 or better in 61.9%. CDVA was 20/40 or better in 100% of eyes and 20/25 or better in 90.2%. Mean refractive cylinder decreased significantly from -2.23 ± 1.72 before to -0.43 ± 0.53 after surgery (P < .05). Vector analysis of attempted versus achieved correction showed 100% of eyes within ± 1.00 D and 95.2% within ± 0.50 D for J and 100% of eyes were within ± 1.00 D and 95.2% within ± 0.50 D for J. Mean toric IOL axis rotation was 4.42 ± 4.31 degrees (range: 0 to 16 degrees) and 86% of the lenses rotated less than 10 degrees.

CONCLUSIONS

The AT TORBI 709 M IOL has proved to be a predictable and effective device with good rotational stability for the correction of preexisting astigmatism during cataract surgery.

摘要

目的

描述 toric AT TORBI 709 M(前 AT.Comfort 464 TLC)人工晶状体(IOL)的眼内旋转稳定性和屈光结果。

方法

纳入 32 例连续白内障患者的 48 只眼,这些患者均存在大于 1 屈光度的角膜散光。在超声乳化白内障吸除术后植入 toric IOL。在术前、术后 1、3 和 6 个月测量未矫正的远视力(UDVA)、矫正的远视力(CDVA)、残余屈光球镜度、残余角膜散光和屈光柱镜度以及 toric IOL 轴位。

结果

术后平均 logMAR UDVA 和 CDVA 显著提高(P <.05)。88.1%的眼 UDVA 达到 20/40 或更好,61.9%的眼达到 20/25 或更好。100%的眼 CDVA 达到 20/40 或更好,90.2%的眼达到 20/25 或更好。平均柱镜度从术前的-2.23 ± 1.72 显著降低至术后的-0.43 ± 0.53(P <.05)。尝试矫正与实际矫正的矢量分析显示,J 轴的 100%的眼在± 1.00 D 内,95.2%的眼在± 0.50 D 内;J 轴的 100%的眼在± 1.00 D 内,95.2%的眼在± 0.50 D 内。平均 toric IOL 轴旋转为 4.42 ± 4.31 度(范围:0 至 16 度),86%的晶状体旋转小于 10 度。

结论

AT TORBI 709 M IOL 已被证明是一种可预测且有效的装置,在白内障手术中矫正术前散光时具有良好的旋转稳定性。