Division of Hematopathology, University of Pittsburgh, Pittsburgh, Pennsylvania 15213-2582, USA.
Cytometry B Clin Cytom. 2013 May;84(3):149-56. doi: 10.1002/cyto.b.21077. Epub 2013 Mar 8.
Laboratories utilize diverse methods to determine the absolute count used to distinguish chronic lymphocytic leukemia (CLL) from monoclonal B lymphocytosis, with many using dual platform (DP) methods (flow cytometric percent of CLL-like B cells × hematology analyzer white blood cell count). However, flow cytometric tube-to-tube variation may make interpretation difficult, particularly when the CLL-like B-cell count straddles the recommended threshold. This study investigates the extent, and potential sources, of this variability.
Flow cytometric enumeration was performed on 20 samples with CLL-like B cells and 10 control specimens using single platform (SP) and two DP tubes A: kappa, lambda, CD5, CD10, CD19, CD38, CD45, CD20 and B: FMC-7, CD23, CD19, CD5, with and without racking to remove doublets. In addition, reproducibility studies were performed.
Three specimens showed discordant CLL-like B-cell counts relative to the threshold of 5 × 10⁹ /L. Bland-Altman analysis demonstrated a systematic bias with DP Tube B > A (5.9%) and SP > DP Tube A (9.5%). The doublet percent varied between specimens, but did not account for the bias between DP tubes. Variation was also seen in enumeration of other cell types, suggesting multiple potential sources of inconsistency.
Significant tube-to-tube variation may be seen in CLL-like B-cell counts. The precise cause of these differences is uncertain, but is likely multifactorial. If the clinical utility of an absolute threshold for the diagnosis of CLL can be confirmed, it will be important to establish recommendations for standardization, similar to those employed for CD4 and CD34 enumeration.
实验室采用多种方法来确定绝对计数,用于区分慢性淋巴细胞白血病(CLL)和单克隆 B 淋巴细胞增多症,许多实验室使用双平台(DP)方法(流式细胞仪检测的 CLL 样 B 细胞百分比×血液分析仪白细胞计数)。然而,流式细胞仪管间的变异可能会使解释变得困难,尤其是当 CLL 样 B 细胞计数跨越推荐阈值时。本研究调查了这种变异性的程度和潜在来源。
使用单平台(SP)和两个 DP 管 A:kappa、lambda、CD5、CD10、CD19、CD38、CD45、CD20 和 B:FMC-7、CD23、CD19、CD5,对 20 个 CLL 样 B 细胞样本和 10 个对照样本进行流式细胞术计数,其中包括有和无排管以去除双联体的情况。此外,还进行了重复性研究。
三个样本相对于 5×10⁹/L 的阈值显示出不一致的 CLL 样 B 细胞计数。Bland-Altman 分析显示 DP 管 B >A(5.9%)和 SP >DP 管 A(9.5%)存在系统偏差。双联体百分比在标本之间存在差异,但不能解释 DP 管之间的偏差。其他细胞类型的计数也存在差异,表明存在多种潜在的不一致来源。
在 CLL 样 B 细胞计数中可能会观察到明显的管间差异。这些差异的确切原因尚不确定,但可能是多因素的。如果可以证实绝对阈值在 CLL 诊断中的临床实用性,那么建立标准化建议将非常重要,类似于 CD4 和 CD34 计数所采用的建议。